Key result
Only 59% of patients with Medtronic ICDs could hear at least one alert tone, and this ability decreased significantly with advancing age (P<0.001).
Why the study?
Are patients with implantable cardioverter-defibrillators able to hear the audible Patient Alert tones?
Cross-Sectional (n=102)
Are patients with implantable cardioverter-defibrillators able to hear the audible Patient Alert tones?
p-value: p=<0.001
A large proportion of ICD patients, particularly older adults, cannot hear the audible Patient Alert tones, suggesting that current guidelines relying on this feature may be inadequate and nonauditory alerts are needed.
Many ICD patients may miss audible alerts, especially older adults; leaves open whether nonauditory alternatives improve safety.
BACKGROUND: Current management guidelines for patients with Medtronic Sprint Fidelis ICD leads (Medtronic Inc., Minneapolis, MN, USA) include prominent use of Patient Alert, a feature in which the ICD generator emits audible beeps at two programmable frequencies. Because hearing loss is highly prevalent beyond the sixth decade of life, the utility of this feature is unclear. Therefore, we conducted a survey of patients' ability to hear the Medtronic Patient Alert. METHODS: During visits to an outpatient device clinic, patients with Medtronic ICDs were evaluated for their ability to hear ICD tones. RESULTS: The patient group consisted of 102 patients. Patients older than 70 years comprised 68% of the sample, with 16% between 60 and 70, and 17% younger than 60 years. Of the 102 patients, 59% (56% of males and 70% of females) were able to hear at least one tone. Ability to hear ICD tones decreased with advancing age. Among patients over 60 and 70 years, 52% and 43%, respectively (P < 0.001 vs. patients below 60 and 70 years), could hear at least one tone. CONCLUSIONS: The Patient Alert feature is not useful among a large proportion of ICD patients. Patients with Sprint Fidelis leads should be evaluated for their ability to hear audible ICD tones. For patients who cannot hear the Patient Alert feature, a wireless remote monitoring and/or daily application of a magnet by a caregiver should be considered. Device manufacturers should include nonauditory alert technologies such as wireless remote monitoring and vibratory stimulation in future devices.
No takes yet. Share an insight, caveat, or question.
SIMONS et al. (2009) conducted a cross-sectional in Patients with Implantable Cardioverter-Defibrillators (ICDs) (n=102). Medtronic Patient Alert tones was evaluated on Ability to hear at least one ICD tone (p=<0.001). Only 59% of patients with Medtronic ICDs could hear at least one alert tone, and this ability decreased significantly with advancing age (P<0.001).