Evaluates measurement methods and the stability of hydrophilic lens dimensions, indicating their importance for accurate fitting.
Most methods of measuring hard lens dimensions cannot be used with hydrophilic lenses. However, practitioners must be able to measure the dimensions of hydrophilic lenses with ease and accuracy. In addition, the lens dimensions must remain relatively stable to provide a proper patient response. This study was designed to evaluate methods of measuring the dimensions of hydrophilic contact lenses and to use these methods to examine the stability of hydrophilic lens dimensions under various conditions. Back surface curvature was measured with a series of templates. Thickness was measured using a modification of the optical method of measuring hard lens thickness. Power was measured on a lensometer after the lens surfaces were blotted. Diameter was measured with a hand comparator, making sure no pressure was placed on the lenses. With these methods, we were able to measure all the dimensions except back surface curvature to within hard lens standards. We found that hydrophilic lens dimensions could be altered significantly by storage in improper solutions, by dehydrating and rehydrating, and by extended boiling.