Externalized conductor cables on recalled Riata ICD leads carry a previously unknown risk of large or calcified thrombus formation, suggesting a need for echocardiographic screening prior to lead extraction.
Answers are grounded in this paper and cite their sources.
Potential thrombus on externalized Riata leads warrants extraction caution; leaves open need for prospective screening studies.
St. Jude Riata family ICD leads are subject to an FDA class I recall due to insulation failure, cable extrusions, and a high rate of electrical failures. We present multiple cases of large intravascular thrombus formation adherent to externalized conductor cables on Riata leads. Our observations highlight a previously unknown risk of large or calcified thrombus formation involving externalized conductors. We suggest that these leads should be screened for thrombi prior to potential lead extraction to determine the safest approach. Serial echocardiography may help detect progressive thrombus formation. The role of systemic anticoagulation for prevention or treatment of these thrombi remains unclear.
No takes yet. Share an insight, caveat, or question.
Goyal et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: