Key result
Riata lead extraction triggers paradoxical thromboembolic stroke via persistent foramen ovale in case report.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Reports paradoxical thromboembolic stroke via persistent foramen ovale during lead extraction, suggesting mandatory transesophageal echocardiography screening.
Recalled St Jude Medical Riata leads have a high rate of insulation failures with potential thrombogenic externalized conductors. We report a paradoxical thromboembolic stroke through a persistent foramen ovale during extraction of a lead with externalization. We suggest mandatory transesophageal echocardiography prior to extraction to screen for lead-related thrombosis and routes of right-to-left shunting to reduce risk of thromboembolic complications.
No takes yet. Share an insight, caveat, or question.
Larsen et al. (2013) conducted a case report in Paradoxical thromboembolic stroke through a persistent foramen ovale during defibrillator lead extraction (n=1). Extraction of recalled St Jude Medical Riata defibrillator lead with conductor externalization was evaluated on Paradoxical thromboembolic stroke. Extraction of a recalled St Jude Medical Riata lead with conductor externalization resulted in 1 case of paradoxical thromboembolic stroke through a persistent foramen ovale.