Why the study?
Serial ostial and distal left main lesions remain technically challenging with potential complications, prompting discussion of stent deformation, explantation risks and benefits, and testing of stent crush behavior.
Does crushing high radial strength stents in the left main lead to inadequate minimal stent area compared to conventional stents?
Does crushing high radial strength stents in the left main lead to inadequate minimal stent area compared to conventional stents?
Bench testing demonstrates that crushing deformed high radial strength stents in the left main leads to inadequate expansion, supporting stent explantation in such scenarios.
Answers are grounded in this paper and cite their sources.
May inform individualized explantation decisions after crushing high radial strength left main stents; leaves open prospective validation of outcomes.
Serial ostial and distal left main lesions continue to be one of the most difficult tasks for the interventional cardiologist, with many potential complications occurring. We present such a high-risk percutaneous coronary intervention where immediate stent explantation was deemed necessary because the metal deformation and high radial strength of the particular stent platform would prevent an acceptable procedural result if it had been crushed to the vessel wall. The aim of this paper was to discuss left main stent deformation, debate the risks and benefits of stent explantation and finally test in-vitro our theory on "insufficient" crush with stents with high radial strength and compare it with conventional stents. Bench-testing supports our ad-hoc explantation decision showing stent underexpansion, recoil, and malapposition, obtaining an inadequate minimal stent area.