Key result
Disabling ventricular automated capture management eliminates short-long-short triggered VF in a CRT-D patient.
Why the study?
Automated capture management (ACM) algorithms are widely used to improve device longevity and reduce manual follow-ups, but their safety and personalization needs remain unclear.
Case Report (n=1)
No
In pacemaker-dependent CRT-D patients, automated ventricular capture management threshold testing can cause transient loss of capture and short-long-short sequences that trigger polymorphic VT/VF, necessitating individualized fixed-output programming.
Answers are grounded in this paper and cite their sources.
May warrant individualized fixed-output programming in CRT-D patients to avoid pacing-induced VF; leaves open proarrhythmic risks of automated capture management.
With rapid advances in pacing and defibrillation systems, various fully automated pacing algorithms are available from different vendors to improve workflow and patient safety1–6. Automated capture management (ACM) increases device longevity and reduces the need for frequent manual device interrogations and follow-ups 6,7. In the current-generation devices, automatic programming includes not only capture threshold testing but also assessment of sensing amplitude, battery voltage, lead impedance, mode switching, rate response, and atrio-ventricular interval adjustments, to name a few.
No takes yet. Share an insight, caveat, or question.
Shrestha et al. (2026) conducted a case report in Pacemaker-dependent CRT-D recipient with recurrent ventricular arrhythmias and Automated Capture Management (ACM)-mediated polymorphic ventricular tachycardia / ventricular fibrillation (n=1). Disabling automated capture management (ACM), reprogramming to fixed pacing outputs with an increased safety margin, and initiation of amiodarone was evaluated on Resolution of ACM-triggered ventricular arrhythmias and ICD shocks. Disabling ventricular automated capture management and reprogramming to fixed pacing outputs eliminated short-long-short triggered ventricular fibrillation and ICD shocks in a CRT-D patient.