The atrial capture management algorithm in CRT-P devices can cause recurrent asystole in pacemaker-dependent patients in the setting of a malfunctioning ventricular lead.
Answers are grounded in this paper and cite their sources.
Warns of ACM algorithm-induced asystole in pacemaker-dependent CRT-P patients; leaves open prospective safety assessments across device platforms.
; Medtronic, Minneapolis, MN, USA) who developed frequent periods of asystole after malfunction of one of the ventricular leads. The underlying cause of asystole was found to be due to the atrial capture management (ACM) algorithm of the CRT-P device, temporarily converting biventricular to right ventricular-only pacing as part of the algorithm. This case highlights implications of the ACM algorithm in devices with a similar platform for pacemaker-dependent patients.