Key result
Asystole due to atrioventricular cross-talk on spontaneous left atrial electrograms was observed in 11% of recipients of ventricular resynchronization stimulation systems.
Why the study?
Does biventricular pacing with LV leads at conventional sensitivity cause ventricular standstill due to atrioventricular cross-talk in patients with congestive heart failure?
Observational (n=17)
Does biventricular pacing with LV leads at conventional sensitivity cause ventricular standstill due to atrioventricular cross-talk in patients with congestive heart failure?
Answers are grounded in this paper and cite their sources.
Atrioventricular cross-talk from LV leads placed in the great cardiac vein can cause ventricular pacing inhibition and potentially fatal asystole in patients with biventricular pacemakers.
The aim of the study was to evaluate, in recipients of biventricular pacing systems, the risk of asystole due to ventricular pacing inhibition by sensing the left atrial signals by the LV lead at conventional sensitivity. Long-term ventricular sensitivity was programmed at > or = 4 m V in 17 consecutive recipients of ventricular resynchronization systems implanted for chronic management of congestive heart failure. Ventricular pacing inhibition due to AV cross-talk on spontaneous left atrial electrogram (AVCSA) was tested at a 2 mV ventricular sensitivity immediately after implantation of the stimulation system and 1 month later. Pacemaker dependence was also tested during temporary VVI pacing at a rate of 30 beats/min. AVCSA was observed in three patients. It was present on the day of implantation in one patient, and developed within the first month in two others. Asystole was observed in two of the three cases of AVCSA. Three pacemaker nondependent patients at the time of system implantation had become pacemaker dependent at 1 month. AVCSA was observed only with LV leads positioned in the great cardiac vein. In conclusion, asystole due to AVCSA was observed in 11% of recipients of ventricular resynchronization stimulation systems. Care should be taken in these patients to minimize the risk of atrial sensing by the LV lead, preferably avoiding its placement in the great cardiac vein. This phenomenon could be eliminated by the programmability of a right ventricular only sensing configuration.
No takes yet. Share an insight, caveat, or question.
Benchaa et al. (2002) conducted an observational in Congestive heart failure (n=17). Biventricular pacing systems was evaluated on Asystole due to atrioventricular cross-talk on spontaneous left atrial electrogram (AVCSA). Asystole due to atrioventricular cross-talk on spontaneous left atrial electrograms was observed in 11% of recipients of ventricular resynchronization stimulation systems.