Why the study?
First-generation leadless pacemakers carried a risk of pacemaker syndrome in sinus rhythm patients, prompting evaluation of a second-generation system providing atrioventricular synchrony via atrial mechanical sensing.
Does a second-generation AV synchronous leadless pacemaker prevent pacemaker syndrome compared to a first-generation VVI leadless pacemaker in patients with sinus rhythm and complete atrioventricular block?
Does a second-generation AV synchronous leadless pacemaker prevent pacemaker syndrome compared to a first-generation VVI leadless pacemaker in patients with sinus rhythm and complete atrioventricular block?
Second-generation AV synchronous leadless pacemakers provide significant atrioventricular synchrony and eliminate the risk of pacemaker syndrome compared to first-generation VVI devices in patients with sinus rhythm and complete AV block.
Answers are grounded in this paper and cite their sources.
AV synchronous leadless pacemakers may lower pacemaker syndrome risk; leaves open need for randomized confirmation before practice change.
INTRODUCTION: transcatheter pacing system (VR; Medtronic) was able to provide single-chamber VVI(R) pacing mode only, with a potential risk of pacemaker syndrome in sinus rhythm patients. A second-generation system (AV) now provides atrioventricular synchrony through atrial mechanical (Am) sensing capability (VDD mode). OBJECTIVE: We sought to compare VR and AV systems in sinus rhythm patients with chronic ventricular pacing (Vp) for complete atrioventricular block. METHODS: All consecutive patients implanted with an LPM in our department for complete atrioventricular block were retrospectively screened. Patients with atrial fibrillation, sinus dysfunction, or Vp burden <20% at 1 month postimplantation were excluded. Patients were systematically followed with a visit at 1 month, and then at least once a year. RESULTS: A total of 93 patients-45 VR (2015-2020) and 48 AV (2020-2021)-were included. VR and AV patients had similar baseline characteristics, except for VR patients being older (80 ± 8 vs. 77 ± 9 years, p = 0.049). The mean Vp burden was 77% in the VR and 82% in the AV group (p = 0.38). In AV patients, the median AV synchronous beats rate was 78%, with 65% having a >66% rate. An E/A ratio <1.2 as measured on echocardiography was the only independent predictor of accurate atrial mechanical tracking (p = 0.01). One-year survival rate was similar in both groups. Five patients in the VR and 0 in the AV group eventually developed pacemaker syndrome within 1 year post-implantation (p = 0.02). CONCLUSION: In sinus rhythm patients with chronic Vp for complete atrioventricular block implanted with an LPM, the atrial mechanical sensing algorithm allowed significant atrioventricular synchrony in most patients and was associated with no occurrence of-otherwise rare-pacemaker syndrome.
No takes yet. Share an insight, caveat, or question.
Lenormand et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: