Why the study?
Does Absorb BVS compared to Mirage BRMS alter the haemodynamic microenvironment (endothelial shear stress) in scaffolded vessels?
Does Absorb BVS compared to Mirage BRMS alter the haemodynamic microenvironment (endothelial shear stress) in scaffolded vessels?
Absorb BVS and Mirage BRMS create different haemodynamic microenvironments, highlighting the potential value of computational modeling in optimizing scaffold configuration.
Answers are grounded in this paper and cite their sources.
Scaffold-specific shear stress patterns in animal models; leaves open computational optimization of BVS geometry for clinical translation.
Absorb demonstrated higher shear stress on top of the struts compared to Mirage. However, in the inter-strut zones shear stress was higher in Mirage than in Absorb. Further research is required to examine the potential value of in vivo computational modelling in optimising scaffold configuration and clinical outcomes.
No takes yet. Share an insight, caveat, or question.
Tenekecioğlu et al. (2017) studied this question.