Key result
Lux-Valve Plus cuts 6-month perivalvular leakage to with first-generation Lux-Valve.
Why the study?
Does TTVR with Lux-Valve Plus improve procedural and echocardiographic outcomes compared to Lux-Valve in patients with severe tricuspid regurgitation?
Cohort (n=28)
No
Does TTVR with Lux-Valve Plus improve procedural and echocardiographic outcomes compared to Lux-Valve in patients with severe tricuspid regurgitation?
Lux-Valve Plus offers greater minimally invasive benefits and lower rates of perivalvular leakage compared to Lux-Valve for transcatheter tricuspid valve replacement.
Answers are grounded in this paper and cite their sources.
Comparable efficacy in this Level 5 study does not support device selection changes; leaves open prospective validation of remodeling durability.
Objective To compare the efficacy of transcatheter tricuspid valve replacement (TTVR) using Lux-Valve and Lux-Valve Plus in patients with severe tricuspid regurgitation. Methods A total of 28 consecutive patients with severe tricuspid regurgitation who underwent TTVR with Lux-Valve (n=14) or Lux-Valve Plus (n=14) in the First Affiliated Hospital of the Air Force Medical University from August 2019 to November 2023 were enrolled. Transthoracic echocardiography was performed in all patients before and 6 months after the TTVR. The ultrasound indexes were compared before and 6 months after the TTVR in all patients and between Lux-Valve and Lux-Valve Plus groups. Results Compared with the Lux-Valve group, the Lux-Valve Plus group showed significantly reduced intraoperative bleeding and shorter postoperative hospital stays (both P<0.05). Six months after the TTVR, none of the patients exhibited more than a mild tricuspid valve regurgitation, and none of the patients had moderate or above perivalvular leakage except for one patient in the Lux-Valve Plus group who had a separation of the clamping member from the anterior tricuspid leaflet. The incidence of perivalvular leakage was significantly lower in the Lux-Valve Plus group (14.29%, 2/14) than in the Lux-Valve group (64.29%, 9/14, P<0.05). At 6 months after operation, the right chamber volume and right ventricle middle transverse diameter were reduced (both P<0.05); the peak blood flow velocity across the tricuspid valve, peak pressure gradient across the tricuspid valve, mean blood flow velocity of tricuspid valve, mean pressure gradient across the tricuspid valve and velocity time integral were increased in both groups (all P<0.05).Compared with the Lux-Valve group, the Lux-Valve Plus group showed higher left ventricular ejection fraction at 6 months postoperatively (P<0.05), while the rest of the indicators were not statistically different (all P>0.05). Conclusions The efficacy of using Lux-Valve and Lux-Valve Plus for TTVR in patients with severe tricuspid regurgitation is comparable. Six months after the TTVR, the right side of the heart has undergone reverse remodeling.While Lux-Valve Plus offers greater minimally invasive benefits, valve selection should consider device-specific characteristics and differences in individual patients.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2025) conducted a cohort in severe tricuspid regurgitation (n=28). Lux-Valve Plus vs. Lux-Valve was evaluated on Incidence of perivalvular leakage at 6 months (p=<0.05). Transcatheter tricuspid valve replacement with Lux-Valve Plus resulted in a lower incidence of perivalvular leakage at 6 months compared to Lux-Valve.