The HemoCue 201(+) showed greater discriminating power for detecting anemia in prospective blood donors than the micro-Hct method. Both presented equivalent deferral error rates of nonanemic potential donors. Compared to the micro-Hct, HemoCue 201(+) reduces the risk of anemic female donors giving blood, specially for those with lower Hb levels, without increasing the deferral of nonanemic potential donors.