Why the study?
Does an AAI-DDD changeover mode (SafeR) reduce ventricular pacing and the composite of hospitalization for heart failure, atrial fibrillation, or cardioversion in patients with sinus node disease or AV block requiring a dual-chamber pacemaker?
Does an AAI-DDD changeover mode (SafeR) reduce ventricular pacing and the composite of hospitalization for heart failure, atrial fibrillation, or cardioversion in patients with sinus node disease or AV block requiring a dual-chamber pacemaker?
The SafeR pacing mode safely reduces right ventricular pacing and may reduce cardiac death or HF hospitalizations, but does not significantly reduce the primary composite of HF hospitalization, AF, or cardioversion compared to standard DDD pacing.
Answers are grounded in this paper and cite their sources.
SafeR fails to reduce the primary composite despite lowering ventricular pacing; extends evidence that pacing minimization is feasible without excess events.
AIM: Right ventricular pacing (VP) has been hypothesized to increase the risk in heart failure (HF) and atrial fibrillation (AF). The ANSWER study evaluated, whether an AAI-DDD changeover mode to minimize VP (SafeR) improves outcome compared with DDD in a general dual-chamber pacemaker population. METHODS AND RESULTS: ANSWER was a randomized controlled multicentre trial assessing SafeR vs. standard DDD in sinus node disease (SND) or AV block (AVB) patients. After a 1-month run-in period, they were randomized (1 : 1) and followed for 3 years. Pre-specified co-primary end-points were VP and the composite of hospitalization for HF, AF, or cardioversion. Pre-specified secondary end-points were cardiac death or HF hospitalizations and cardiovascular hospitalizations. ANSWER enrolled 650 patients (52.0% SND, 48% AVB) at 43 European centres and randomized in SafeR (n = 314) or DDD (n = 318). The SafeR mode showed a significant decrease in VP compared with DDD (11.5 vs. 93.6%, P < 0.0001 at 3 years). Deaths and syncope did not differ between randomization arms. No significant difference between groups [HR = 0.78; 95% CI (0.48-1.25); P = 0.30] was found in the time to event of the co-primary composite of hospitalization for HF, AF, or cardioversion, nor in the individual components. SafeR showed a 51% risk reduction (RR) in experiencing cardiac death or HF hospitalization [HR = 0.49; 95% CI (0.27-0.90); P = 0.02] and 30% RR in experiencing cardiovascular hospitalizations [HR = 0.70; 95% CI (0.49-1.00); P = 0.05]. CONCLUSION: SafeR safely and significantly reduced VP in a general pacemaker population though had no effect on hospitalization for HF, AF, or cardioversion, when compared with DDD.
No takes yet. Share an insight, caveat, or question.
Stockburger et al. (2014) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: