Can chest compressions during CPR cause inappropriate shock delivery from an S-ICD due to oversensing?
Can chest compressions during CPR cause inappropriate shock delivery from an S-ICD due to oversensing?
Chest compressions during CPR can cause S-ICD oversensing leading to inappropriate shock delivery, highlighting a critical safety consideration during resuscitation.
Answers are grounded in this paper and cite their sources.
May prompt CPR protocol caution with S-ICDs; leaves open whether sensing adjustments reduce inappropriate shocks.
We report the case of a patient who received both appropriate and inappropriate shocks from an entirely subcutaneous implantable defibrillator (S-ICD). The inappropriate shocks were due to oversensing of chest compressions in the setting of a profound postshock bradycardia. It is important to recognize that the therapy from the S-ICD can not only be withheld, as previously described, but can also be inappropriately delivered during chest compressions.