Why the study?
Is a nonthoracotomy ICD system using a single subcutaneous array lead and abdominal active can safe and effective for preventing sudden cardiac death in a small child with HCM?
Is a nonthoracotomy ICD system using a single subcutaneous array lead and abdominal active can safe and effective for preventing sudden cardiac death in a small child with HCM?
Answers are grounded in this paper and cite their sources.
A nonthoracotomy ICD system using a single subcutaneous array lead and abdominal active can appears safe and feasible for preventing sudden cardiac death in small children.
This report describes the case of an 8-year-old boy with hypertrophic cardiomyopathy (HCM) who underwent ICD implantation for recurrent syncope. To avoid vascular complications and to minimize the surgical approach in this small child, a nonthoracotomy ICD system was chosen using a single subcutaneous array lead with only one finger, an abdominally placed active can, and epicardial dual chamber pacing and sensing electrodes. During an 8-month follow-up, DFT was confirmed and there were no ventricular tachycardia or complications. It appears to be a safe device for the prevention of sudden cardiac death in infants and small children.
No takes yet. Share an insight, caveat, or question.
Luedemann et al. (2004) studied this question.