Why the study?
Current voltage mapping to identify slow conducting channels during scar-related VA ablation is time-consuming, requiring manual electrogram analysis, and limited by electric far-field.
Does the SCC-Mapping algorithm improve the identification of slow conducting channels compared to standard voltage mapping in patients undergoing ventricular arrhythmia ablation?
Does the SCC-Mapping algorithm improve the identification of slow conducting channels compared to standard voltage mapping in patients undergoing ventricular arrhythmia ablation?
An automated SCC-Mapping algorithm provides operator-independent and superior identification of arrhythmogenic substrates compared to standard voltage mapping during ventricular arrhythmia ablation.
Answers are grounded in this paper and cite their sources.
May aid substrate identification in VA ablation; leaves open whether it improves outcomes versus standard EAM.
BACKGROUND: Voltage mapping allows identifying the arrhythmogenic substrate during scar-related ventricular arrhythmia (VA) ablation procedures. Slow conducting channels (SCCs), defined by the presence of electrogram (EGM) signals with delayed components (EGM-DC), are responsible for sustaining VAs and constitute potential ablation targets. However, voltage mapping, as it is currently performed, is time-consuming, requiring a manual analysis of all EGMs to detect SCCs, and its accuracy is limited by electric far-field. We sought to evaluate an algorithm that automatically identifies EGM-DC, classifies mapping points, and creates new voltage maps, named "Slow Conducting Channel Maps" (SCC-Maps). METHODS: Retrospective analysis of electroanatomic maps (EAM) from 20 patients (10 ischemic, 10 with arrhythmogenic right ventricular dysplasia/cardiomyopathy) was performed. EAM voltage maps were acquired during sinus rhythm and used for ablation. Preprocedural contrast-enhanced cardiac magnetic resonance (Ce-CMR) imaging was available for the ischemic population. Three mapping modalities were analysed: (i) EAM voltage maps using standard (EAM standard) or manual (EAM screening) thresholds for defining core and border zones; (ii) SCC-Maps derived from the use of the novel SCC-Mapping algorithm that automatically identify EGM-DCs measuring the voltage of the local component; and (iii) Ce-CMR maps (when available). The ability of each mapping modality in identifying SCCs and their agreement was evaluated. CONCLUSION: The SCC-Mapping algorithm allows an operator-independent analysis of EGM signals showing better identification of the arrhythmogenic substrate characteristics when compared to standard voltage EAM.
No takes yet. Share an insight, caveat, or question.
Alcaine et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: