Why the study?
Lead failure is the major limitation in ICD therapy, and long-term follow-up data for Biotronik Linox ICD leads are limited.
Does the performance and survival of Biotronik Linox ICD leads differ from Linox Smart ICD leads?
Does the performance and survival of Biotronik Linox ICD leads differ from Linox Smart ICD leads?
Answers are grounded in this paper and cite their sources.
Biotronik Linox and Linox Smart ICD leads have comparable performance, but show a relevant failure rate with a marked drop in survival probability after 5 years of follow-up.
BACKGROUND: Lead failure is the major limitation in implantable cardioverter-defibrillator (ICD) therapy. Long-term follow-up data for Biotronik Linox ICD leads are limited. Therefore, we analyzed the performance of all these leads implanted at our institution. MATERIALS AND METHODS: All Linox and Linox Smart ICD leads implanted between 2006 and 2015 were identified. Lead failure was defined as electrical dysfunction (oversensing, abnormal impedance, exit block). Lead survival was described, according to Kaplan-Meier. Associations between lead failure and specific variables were examined. p < .05 was considered significant. RESULTS: We included 417 ICD leads. The median follow-up time for Linox (n = 205) was 81 months and for Linox Smart (n = 212) 75 months. During that follow-up time, 30 Linox (14.6%) and 16 Linox Smart leads (7.6%) showed a malfunction. The 5-year lead survival probability was 97.4% for Linox and 95.2% for Linox Smart (log-rank test, p = .19). The 6- and 8-year lead survival probability for Linox was 93.6% and 84.6%, and for Linox Smart 93% and 91.9%. The only factor significantly associated with lead failure was younger patient age at implantation (hazard ratio/year: 0.97, 95% CI: 0.94-0.99, p = .002). CONCLUSION: This relatively large study with a long follow-up period highlights a relevant failure rate of Biotronik Linox leads. The performance of Linox versus Linox Smart ICD leads was comparable. Although we show an acceptable 5-year lead survival probability, we observed a marked drop after just 1 more year of follow-up. In an era of improving heart failure survival probability, a prolonged follow-up of ICD leads is increasingly clinically relevant.
No takes yet. Share an insight, caveat, or question.
Klampfleitner et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: