Key result
Linox ICD leads linked to ~8-fold higher failure risk versus Durata, especially in women.
Why the study?
Does the Linox ICD lead have a higher rate of failure compared to the Durata ICD lead in patients receiving ICDs?
Cohort (n=1,315)
Yes
Does the Linox ICD lead have a higher rate of failure compared to the Durata ICD lead in patients receiving ICDs?
Absolute Event Rate: 3.4% vs 0.4%
p-value: p=< 0.001
The Biotronik Linox ICD lead exhibits a significantly higher failure rate compared to the Durata lead, particularly in female patients, warranting ongoing surveillance.
May warrant closer Linox lead surveillance in women; leaves open whether design or selection factors explain excess failures.
INTRODUCTION: The Linox and Durata implantable cardioverter defibrillator (ICD) leads were introduced to British Columbia (BC) in 2008. We determined their performance and the potential risk factors for lead failure in a large population-based patient registry. METHODS AND RESULTS: We used the BC Cardiac Registry, a mandatory Governmental database of ICD implants, to identify all recipients of Linox and Durata leads in BC between October 2008 and April 2012, and those subsequently undergoing reoperation. Lead failure was defined as recurrent nonphysiological high-rate sensing unrelated to external electromagnetic interference or T-wave oversensing; a sudden rise in impedance unrelated to perforation or lead dislodgement; or abnormal lead parameters with definite evidence of lead fracture or insulation failure. We determined the estimated cumulative lead survival by the Kaplan-Meier method, and the risk factors for lead failure in a proportional hazards model. Over a median of 39 (27-50) months, the Linox failed more frequently than the Durata (16/477 [3.4%] vs. 4/838 [0.4%]; P < 0.001), and had a significantly lower estimated cumulative survival (91.6 [80.6-96.5]% vs. 99.4 [98.4-99.8]% at 5 years; P < 0.0001). Linox failure was characterized by high-rate nonphysiological sensing (11 cases), and/or a sudden impedance rise (7 cases). Insulation failure was clearly confirmed in 6 cases of Linox failure. Female sex was a significant risk factor for Linox failure (adjusted HR = 2.1[1.3-3.4]; P = 0.004). CONCLUSIONS: This multicenter registry indicates a high rate of Linox lead failure, particularly in female patients. Ongoing surveillance of the Linox ICD lead performance is recommended.
No takes yet. Share an insight, caveat, or question.
Padfield et al. (2014) conducted a cohort in Implantable cardioverter defibrillator (ICD) recipients (n=1,315). Linox ICD lead vs. Durata ICD lead was evaluated on Lead failure (p=< 0.001). The Linox ICD lead had a significantly higher failure rate than the Durata lead (3.4% vs 0.4%; P<0.001), with female sex identified as a significant risk factor for Linox failure.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: