Key result
Linox S DX leads maintain stable atrial sensing but misclassify ~40% of arrhythmic episodes.
Why the study?
Single-electrode ICDs with a floating atrial dipole were introduced to overcome the risk of inappropriate shocks, and this study evaluated the long-term performance of a new generation lead (Linox S DX).
What is the long-term sensing performance and arrhythmia classification accuracy of a single-chamber ICD lead with a floating atrial dipole in patients requiring an ICD?
Cohort (n=93)
What is the long-term sensing performance and arrhythmia classification accuracy of a single-chamber ICD lead with a floating atrial dipole in patients requiring an ICD?
While the single-lead ICD with a floating atrial dipole provides stable long-term atrial sensing, device software misclassified 40.2% of arrhythmic episodes, resulting in a 7.5% inappropriate therapy rate and missed ventricular tachycardia events in three patients.
Answers are grounded in this paper and cite their sources.
Stable atrial sensing may not prevent misclassified episodes or missed VT; leaves open need for software refinements in larger cohorts.
OBJECTIVES: The study aim is to present the long-term performance of a new generation implantable cardioverter defibrillator (ICD) electrode with floating atrial dipole (Linox S DX, Biotronik, Berlin, Germany). BACKGROUND: The single ICD electrode with a floating dipole in the atrial chamber was introduced about 15 years ago to overcome risk of inappropriate shock. METHODS: After implantation, internal electrocardiogram data were prospectively collected via telemonitoring (Home Monitoring, Biotronik). RESULTS: A total of 93 patients (81.5% male, 18.5% female; 58.9 ± 12.3 years) were implanted with a single-chamber ICD using the Linox S DX. Patients were followed up for a median of 693 days (33-2,460 days). At time of implantation average p-wave value was 3.5 ± 1.7 mV and remained stable throughout follow-up with an average value of 3.7 ± 1 mV (P = 0.2). A total of 460 arrhythmic episodes were recorded and 185 (40.2%) were incorrectly stratified by the device software. Seven patients (7.5%) experienced inappropriate ICD therapy. In three patients, VT episodes were not detected and remained untreated. CONCLUSIONS: Although the Linox S DX lead presents a satisfactory long-term stability of the atrial sensing, many of the messages sent to the device during follow-up were incorrectly classified by the ICD software. As a result, inappropriate therapy occurred with a rate similar to that observed with dual-chamber ICDs, but some malignant arrhythmias remained undiagnosed and untreated while occurring. Results in larger prospective cohorts should be analyzed and software improvements of the device should be suggested to overcome these potential drawbacks.
No takes yet. Share an insight, caveat, or question.
Şafak et al. (2017) conducted a cohort in Patients requiring implantable cardioverter defibrillator (ICD) implantation (n=93). Single-chamber ICD with floating atrial sensing dipole (Linox S DX) was evaluated on Long-term performance including atrial sensing stability, incorrect arrhythmia stratification, and inappropriate ICD therapy (p=0.2). The Linox S DX ICD lead maintained stable atrial sensing, but 40.2% of arrhythmic episodes were misclassified and 7.5% of patients had inappropriate therapy.