Key result
Lead Integrity Alert algorithm predicts Biotronik Linox ICD lead failure with ~95% positive predictive value.
Why the study?
The utility of Medtronic's Lead Integrity Alert software algorithm in assessing performance of Biotronik Linox high-voltage ICD leads was unknown.
Does Medtronic's Lead Integrity Alert (LIA) software accurately detect lead failure in patients with Biotronik Linox ICD leads?
Cohort (n=208)
Single-blind
Does Medtronic's Lead Integrity Alert (LIA) software accurately detect lead failure in patients with Biotronik Linox ICD leads?
Medtronic's Lead Integrity Alert (LIA) software is highly predictive of lead failure in Biotronik Linox ICD leads, demonstrating its utility for ongoing surveillance.
Answers are grounded in this paper and cite their sources.
May aid Biotronik Linox lead surveillance with LIA; extends prior data but leaves prospective validation open.
INTRODUCTION: Medtronic's Lead Integrity Alert (LIA) software algorithm is useful for detecting abnormal parameters across various ICD-lead families. However, its utility in the assessment of the Biotronik Linox™ family of high-voltage (HV) leads is unknown. METHODS: We conducted a retrospective cohort study to assess the performance of the LIA algorithm to detect abnormalities and lead failure in Linox ICD-leads. All LIA-enabled Medtronic devices connected to an active Linox lead were included. The alerts were adjudicated by 2 blinded electrophysiologists and correlated with clinical data. RESULTS: Between 2008 and 2012, data from 208 patients with 564 patient-years of follow-up were available for analysis. The median follow-up duration was 32 (IQR 21-41 months). Twenty-one LIA triggers were noted in 20 different patients. The median delay until a positive LIA was 32 months (IQR 21-41 months) postimplant with a 5-year lead survival free from LIA of 76%. Ninety-five percent (19/20) LIA alerts were true lead failures. The most common LIA triggers were short V-V intervals (85%) and nonsustained ventricular tachycardia (85%). Abrupt changes of the ICD-lead impedance occurred in 5/20 triggers. Inappropriate ICD-shocks were strongly associated with a positive LIA (30% vs. 7.4%; P = 0.006). Of the explanted Linox leads 53% had visible abnormalities. The sensitivity, specificity, and positive predictive value for lead failure in the presence of a LIA trigger were 87%, 99.5%, and 95.2%, respectively. CONCLUSIONS: A positive LIA trigger in Biotronik Linox ICD-leads is highly predictive of lead failure. LIA is useful in ongoing surveillance of lead performance.
No takes yet. Share an insight, caveat, or question.
Steinberg et al. (2015) conducted a cohort in Patients with LIA-enabled Medtronic devices and active Linox leads (n=208). Lead Integrity Alert (LIA) software algorithm was evaluated on lead failure. The Lead Integrity Alert algorithm in Biotronik Linox ICD-leads is highly predictive of lead failure, with 87% sensitivity, 99.5% specificity, and 95.2% positive predictive value.
Synapse has enriched 10 related papers. Consider them for comparative context: