Key result
Linox ICD leads linked to high premature failure versus other leads, showing ~76% eight-year survival.
Why the study?
Do Linox ICD leads have higher failure rates compared to other ICD leads in patients undergoing ICD implantation?
Cohort (n=287)
No
Do Linox ICD leads have higher failure rates compared to other ICD leads in patients undergoing ICD implantation?
Linox ICD leads demonstrate significantly higher premature failure rates compared to other contemporary leads, particularly in younger patients.
Answers are grounded in this paper and cite their sources.
Warrants close monitoring of Linox leads; observational data leaves open need for prospective confirmation before practice change.
INTRODUCTION: ICD lead failure is a potential source of significant morbidity and mortality. This study investigates the survival rates of Sprint Quattro, Endotak Reliance, and Linox ICD leads. METHODS AND RESULTS: A retrospective cohort study of all patients with an ICD implanted between January 2007 and December 2012 from the Wellington Hospital region, New Zealand, a tertiary referral center. Lead and patient details were established by review of clinical notes and the PaceArt Optima database. We analyzed a total of 287 implants using Sprint Quattro (n = 92), Endotak Reliance (n = 37), Linox (n = 151), Riata (n = 4), and Sprint Fidelis (n = 2) leads. Median follow-up was 61.6 (26.0-81.6) months for Sprint Quattro leads, 66.7 (53.3-88.8) months for Linox leads, and 82.9 (45.9-107.4) months for Endotak Reliance leads. There were 20 cases of lead failure of which 19 were in Linox leads. The 4-, 6-, and 8-year survival for Linox leads was 97% (92.6-99.1), 93% (85.5-96.5), and 76% (62.3-85.5), respectively. The predominant abnormality was detection of nonphysiological electrical signals. Linox lead failure was associated with a younger age of patient (49.2 vs. 57.7 years, P = 0.007). CONCLUSIONS: The 7-year survival of Linox leads was significantly worse than shown in Biotronik surveillance reports, but in line with other single-center studies from around the world.
No takes yet. Share an insight, caveat, or question.
O’Connor et al. (2018) conducted a cohort in ICD implantation (n=287). Linox ICD leads vs. Sprint Quattro, Endotak Reliance, and other ICD leads was evaluated on lead failure. Linox ICD leads exhibited a high rate of premature failure, accounting for 19 of 20 total failures among 287 implants, with an 8-year lead survival of 76%.