Why the study?
Is Doppler ultrasonic estimation of cardiac output reproducible and accurate compared to standard methods in humans?
Is Doppler ultrasonic estimation of cardiac output reproducible and accurate compared to standard methods in humans?
Doppler ultrasound is a safe, fairly reproducible, and reasonably accurate method for measuring cardiac output in selected patients provided signal quality is adequate.
Answers are grounded in this paper and cite their sources.
Warrants caution with single Doppler aortic measurements clinically; leaves open standardization efforts in research.
The Doppler ultrasonic estimation of cardiac output in man is reviewed. Minimal requirements for accurate measurements are discussed, and the published results of reproducibility studies and validation studies are summarized and analysed. Analysis of Doppler records has a coefficient of repeat determination of 5-8% for aortic or LV outflow tract measurements and this is higher for other sites. Short-term variability varies from 4 to 10%, and that over days to weeks from 9 to 14%. Thus a single measurement may vary up to +/- 28% over time with no true change in cardiac output. For cardiac output determination, the Doppler methods show accuracies varying from 10 to 22% (coefficient of variation of the differences between methods) indicating that a single aortic based measurement only reliably lies within +/- 28% compared with other 'standard' methods, and during exercise the accuracy is less (+/- 44%). Doppler methods are safe, fairly reproducible and reasonably accurate methods for measuring cardiac output in selected patients provided signal quality is adequate during recording.
No takes yet. Share an insight, caveat, or question.
Andrew J.S. Coats (1990) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: