Why the study?
Does primary prevention ICD generator change improve survival in patients with heart failure compared to patients without ICDs?
Does primary prevention ICD generator change improve survival in patients with heart failure compared to patients without ICDs?
Primary prevention ICD generator replacement in patients whose LVEF has improved to >35% or who have LVEF ≤35% with low predicted risk of arrhythmic death is associated with worse survival compared to matched controls without ICDs.
Answers are grounded in this paper and cite their sources.
Generator changes in recovered LVEF>35% were associated with worse survival; leaves open net benefit versus no ICD in low-risk patients.
<0.0001). Patients undergoing GCs of initial noncardiac resynchronization therapy PP ICDs with LVEF >35% had a lower Seattle Heart Failure Model-adjusted survival versus 23 472 control heart failure patients without ICDs (model interaction hazard ratio, 1.21 [95% CI, 1.11-1.31]). In patients undergoing GCs of initial noncardiac resynchonization therapy PP ICDs with LVEF ≤35%, the model indicated worse survival versus controls in the 21% of patients with a PRAD <43% and improved survival in the 10% with PRAD >65%. The association of the PRAD with survival benefit or harm was similar in patients with or without pre-GC ICD therapies. Conclusions Patients who received replacement of an ICD originally implanted for primary prevention and had at the time of GC either LVEF >35% alone or both LVEF ≤35% and PRAD <43% had worse survival versus controls without ICDs.
No takes yet. Share an insight, caveat, or question.
Bilchick et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: