Key result
Inappropriate ICD/CRT-D shocks linked to ~6-fold higher hospitalization rates but not composite adverse outcomes.
Why the study?
Inappropriate shocks from ICDs and CRT-Ds are thought to negatively impact quality of life and prognosis, but differences in outcomes between patients with and without them needed evaluation.
Do inappropriate shocks increase the risk of adverse outcomes in patients with ICD/CRT-D?
Observational (n=220)
Do inappropriate shocks increase the risk of adverse outcomes in patients with ICD/CRT-D?
Absolute Event Rate: 33.3% vs 5.6%
p-value: p=0.016
Inappropriate shocks in ICD/CRT-D patients were not associated with adverse composite outcomes, though they were linked to higher hospitalization rates, with younger age and atrial fibrillation identified as predictors.
Inappropriate shocks may worsen prognosis in ICD/CRT-D patients; leaves open whether prevention strategies improve survival.
Introduction While implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) are effective in preventing sudden death, they are associated with complications, including inappropriate shocks, which have been thought to negatively impact quality of life and prognosis. Objectives This study aimed to evaluate differences in outcomes between patients with and without inappropriate shocks and identify potential predictors of inappropriate shock occurrence. Methods A retrospective analysis between 2020 and 2024 included 264 patients who underwent ICD/CRT-D implantation. Data collected encompassed demographic characteristics, cardiovascular risk factors, documented arrhythmias, medical therapy, etiology, prevention type, and device mode. Outcomes analyzed were hospitalization due to heart failure, myocardial infarction, stroke, or death. Predictors of inappropriate shocks were identified using multivariate logistic regression. Results Of the 264 patients, data on inappropriate shocks were available for 220, with 4% experiencing inappropriate shocks. The sample comprised predominantly men (78.8%) with a median age of 66 (58–73) years. Characteristics between the two groups (with vs. without inappropriate shocks) were largely similar, including device type, with the exception of higher hospitalization rates in the group of patients with inappropriate shocks (33.3% vs. 5.6%, p=0.016). Composite outcomes showed no significant differences between groups. Predictors of inappropriate shocks included age (OR=0.940, 95% CI: 0.886–0.997, p=0.039) and atrial fibrillation (OR=5.279, 95% CI: 1.021–27.286, p=0.047). Conclusion Despite the psychological and economic burden on the healthcare system, inappropriate shocks were not associated with adverse outcomes in this study. However, younger age and atrial fibrillation were identified as predictors of inappropriate shocks. These findings underscore the importance of aggressive atrial fibrillation management (rhythm/heart rate control) and careful optimization of device programming, particularly in younger patients.
No takes yet. Share an insight, caveat, or question.
Pereira et al. (2026) conducted an observational in ICD/CRT-D patients (n=220). Inappropriate shocks vs. No inappropriate shocks was evaluated on Hospitalization (p=0.016). Inappropriate shocks in ICD/CRT-D patients were associated with higher hospitalization rates (33.3% vs. 5.6%, p=0.016) but not composite adverse outcomes, with younger age and AF as key predictors.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: