Key result
Pre-procedural saline bolus linked to ~1.5 mm larger LAA landing zone, improving device sizing agreement.
Why the study?
The safety and effects on LAA and device sizing of administering a fluid bolus during pre-procedural TEE were unknown.
Does a 500 mL intravenous fluid bolus during pre-procedural TEE improve LAA sizing accuracy and predict final device size in patients undergoing LAAC?
Observational (n=72)
No
Does a 500 mL intravenous fluid bolus during pre-procedural TEE improve LAA sizing accuracy and predict final device size in patients undergoing LAAC?
Administering a 500 mL intravenous saline bolus during pre-procedural TEE is safe, significantly increases LAA dimensions, and improves the accuracy of device sizing for LAA closure.
Answers are grounded in this paper and cite their sources.
Volume loading during TEE may improve LAA sizing accuracy; leaves open need for randomized trials before practice change.
BACKGROUND: In patients undergoing left atrial appendage (LAA) closure, an accurate sizing of the LAA is key to optimize device sizing, procedural success and reduce complications. Previous studies have shown that intraprocedural volume loading increases LAA dimensions and improves device sizing. However, the safety and effects on LAA and device sizing of administering a fluid bolus during pre-procedural transesophageal echocardiography (TEE) are unknown. The aim of this study was to determine the safety and impact on LAA dimensions and device sizing of an intravenous (IV) fluid bolus administered during TEE in the setting of the pre-procedural work-up for LAA closure. METHODS: The study included a total of 72 patients who underwent TEE to assess suitability for LAAC and received a 500 ml IV bolus of normal saline. The LAA landing zone (LZ) and depth were measured by TEE before and after volume loading, and these measurements were used to predict the device size implanted during a subsequent percutaneous LAAC procedure. Sizing based on post-bolus measurements of the LZ significantly improved the agreement with the final device size selection during the procedure in 71.0% of cases (vs. 42.0% with pre-bolus measurements). CONCLUSIONS: Volume loading during ambulatory TEE as part of the pre-procedural work-up of LAAC is safe and significantly increases LAA dimensions. This strategy may become the new standard, particularly in centers performing LAAC with no TEE guidance, as it improves LAA sizing and more accurately predicts the final device size.
No takes yet. Share an insight, caveat, or question.
Freitas‐Ferraz et al. (2021) conducted an observational in Atrial fibrillation undergoing pre-procedural work-up for left atrial appendage closure (n=72). Intravenous normal saline bolus vs. Pre-bolus baseline was evaluated on Maximum diameter of the left atrial appendage landing zone at 90 degrees (mm) (p=<0.001). A 500 mL intravenous normal saline bolus administered during pre-procedural transesophageal echocardiography significantly increased the left atrial appendage landing zone diameter by 1.5 mm and improved device sizing agreement from 42.0% to 71.0%.
Synapse has enriched 10 related papers. Consider them for comparative context: