Key result
Implantation of the Tensum silicon-carbide-coated coronary stent was safe and efficacious, with 97% survival and 15% target-vessel revascularization at a mean follow-up of 454 days.
Why the study?
Is the implantation of an amorphous hydrogenated silicon-carbide-coated stent safe and efficacious in patients requiring coronary stenting?
Observational (n=174)
Is the implantation of an amorphous hydrogenated silicon-carbide-coated stent safe and efficacious in patients requiring coronary stenting?
The silicon-carbide-coated Tensum stent demonstrated high primary success and favorable long-term safety and efficacy in daily practice.
Supports real-world use of silicon-carbide-coated stents; leaves open need for randomized comparisons.
The hemocompatibility and biocompatibility of a stent are determined by the physical and electrochemical properties of the stent surface. The aim of this study was to determine the feasibility, safety and efficacy of implantation of a stent coated with silicon carbide. Baseline characteristics were collected prospectively. The occurrence of cardiac adverse events and the angina score were assessed at clinical follow-up. A total of 193 Tensum stents were implanted in 174 patients. In hospital, one patient experienced stent thrombosis and in 6% of the patients a creatinine kinase elevation to 240 U/l or more occurred. Long-term follow-up was performed in 172 patients, with a mean follow-up of 454 +/- 181 days. Ninety-seven per cent were still alive, 15% had undergone target-vessel revascularization, and 2% had angiographic restenosis and were treated with medication only. Seventy-one per cent of the patients were free of anginal complaints, and 20% had anginal complaints in Canadian Cardiac Society class I or II. The Tensum coronary stent showed to be a safe and efficacious device in this study, with a high primary success rate and favorable long-term clinical followup.
No takes yet. Share an insight, caveat, or question.
Hanekamp et al. (1998) conducted an observational in Coronary artery disease (n=174). Tensum stent (amorphous hydrogenated silicon-carbide-coated stent) was evaluated on Cardiac adverse events and angina score. Implantation of the Tensum silicon-carbide-coated coronary stent was safe and efficacious, with 97% survival and 15% target-vessel revascularization at a mean follow-up of 454 days.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: