Key result
Real-time x-ray fused with MRI (XFM) accurately targeted endomyocardial injections in infarcted swine hearts, with a mean distance of 3.2±2.6 mm between predicted and actual locations.
Why the study?
Does real-time x-ray fused with MRI (XFM) accurately guide endomyocardial injections in a swine model of myocardial infarction?
Does real-time x-ray fused with MRI (XFM) accurately guide endomyocardial injections in a swine model of myocardial infarction?
Real-time x-ray fused with MRI (XFM) accurately targets endomyocardial injections in a swine model of myocardial infarction, demonstrating the feasibility of combining 3D MRI soft tissue contrast with high-resolution 2D x-ray fluoroscopy.
XFM accurately targets endomyocardial injections in swine MI; leaves open translation to human use.
BACKGROUND: Magnetic resonance imaging (MRI) permits 3-dimensional (3D) cardiac imaging with high soft tissue contrast. X-ray fluoroscopy provides high-resolution, 2-dimensional (2D) projection imaging. We have developed real-time x-ray fused with MRI (XFM) to guide invasive procedures that combines the best features of both imaging modalities. We tested the accuracy of XFM using external fiducial markers to guide endomyocardial cell injections in infarcted swine hearts. METHODS AND RESULTS: Endomyocardial injections of iron-labeled mesenchymal stromal cells admixed with tissue dye were performed in previously infarcted hearts of 12 Yucatan miniswine (weight, 33 to 67 kg). Features from cardiac MRI were displayed combined with x-ray in real time to guide injections. During 130 injections, operators were provided with 3D surfaces of endocardium, epicardium, myocardial wall thickness (range, 2.6 to 17.7 mm), and infarct registered with live x-ray images to facilitate device navigation and choice of injection location. XFM-guided injections were compared with postinjection MRI and with necropsy specimens obtained 24 hours later. Visual inspection of the pattern of dye staining on 2,3,5-triphenyltetrazolium chloride-stained heart slices agreed (kappa=0.69) with XFM-derived injection locations mapped onto delayed hyperenhancement MRI and the susceptibility artifacts seen on the postinjection T2*-weighted gradient echo MRI. The distance between the predicted and actual injection locations in vivo was 3.2+/-2.6 mm (n=64), and 75% of injections were within 4.1 mm of the predicted location. CONCLUSIONS: Three-dimensional to two-dimensional registration of x-ray and MR images with the use of external fiducial markers accurately targets endomyocardial injection in a swine model of myocardial infarction.
No takes yet. Share an insight, caveat, or question.
Silva et al. (2006) studied Myocardial infarction (n=12). Real-time x-ray fused with MRI (XFM) vs. Postinjection MRI and necropsy was evaluated on Distance between predicted and actual injection locations in vivo. Real-time x-ray fused with MRI (XFM) accurately targeted endomyocardial injections in infarcted swine hearts, with a mean distance of 3.2±2.6 mm between predicted and actual locations.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: