Why the study?
Does a newly developed subvalvular mitral valve repair device reduce regurgitant volume in an in vitro model of functional ischemic mitral regurgitation?
Does a newly developed subvalvular mitral valve repair device reduce regurgitant volume in an in vitro model of functional ischemic mitral regurgitation?
A novel subvalvular mitral valve repair device significantly reduced regurgitant volume in an in vitro porcine model of functional ischemic mitral regurgitation, providing proof-of-concept for its potential clinical utility.
Reduces regurgitant volume in porcine in vitro model; leaves open clinical translation for functional ischemic mitral regurgitation.
BACKGROUND: Optimal surgical treatment of functional ischemic mitral regurgitation (FIMR) is still controversy. Due to the underlying pathophysiology, stand-alone ring annuloplasty is assumed with a high recurrence rate of mitral regurgitation, thus additional subvalvular repair techniques might improve the results. This in vitro study introduces a new device for subvalvular mitral valve repair. METHODS: We developed a new device for subvalvular mitral valve repair, consisting of two frames for papillary muscle (PM) attachment, which are connected with two holding bars serving for fixation of the device on an annuloplasty ring. In the first experimental run, porcine mitral valves including the chordae tendineae and PMs were fixated on a holding device, consisting of a holding ring simulating mitral annulus dilation and height-adjustable frames for PM attachment simulating leaflet tethering. In vitro regurgitant volume was determined in a pulse duplicator. Afterward, the frames for PM attachment were replaced by our newly developed device and the measurements were repeated. CONCLUSION: In this specific in vitro model, our new device for subvalvular mitral valve reconstruction led to a significant reduction of the regurgitant volume, thus representing a promising technique to potentially improve the results of mitral reconstruction in ischemic functional mitral valve regurgitation. Additional studies are required to further investigate and improve our device.
No takes yet. Share an insight, caveat, or question.
Stock et al. (2018) studied Functional Ischemic Mitral Regurgitation. New Device was evaluated.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: