Key result
ChoiceMMed MD300C29 proves the most accurate OTC pulse oximeter, though all lose accuracy in hypoxemic patients.
Why the study?
Over-the-counter pulse oximeters are commonly used in patient care but had not yet been evaluated for their accuracy and precision.
Do over-the-counter pulse oximeter finger devices accurately and precisely measure oxygen saturation compared to arterial blood gas in adult patients?
Population
107 participants aged 18 and above
Comparison
Seven over-the-counter pulse oximeter finger devices vs arterial blood gas SaO2
Design
Cross-sectional study
Authors
Loading...
Caution against unverified OTC oximeter use in hypoxemia; leaves open need for prospective device validation.
Cross-Sectional (n=107)
No
Do over-the-counter pulse oximeter finger devices accurately and precisely measure oxygen saturation compared to arterial blood gas in adult patients?
Over-the-counter pulse oximeters vary significantly in accuracy and precision, with performance notably degrading during hypoxemia, cautioning against their unverified clinical reliance.
Sungahid et al. (2025) conducted a cross-sectional in Patients requiring arterial blood gas testing (n=107). Seven over-the-counter pulse oximeter finger devices vs. Arterial blood gas oxygen saturation (SaO2) was evaluated on Accuracy measured by average root mean square error (ARMS). Among seven over-the-counter pulse oximeters, ChoiceMMed MD300C29 was the most accurate with an average root mean square error of 1.89%, though all devices showed decreased precision and accuracy in hypoxemic patients.