This is the authors' abstract. We don't add key points for this paper.
OBJECTIVE: To determine the correlation between 3 lightweight portable pulse oximeter devices compared to a standard wall mount pulse oximetry device. METHODS: We performed a single-center, prospective, observational study of 4 pulse oximetry devices, 3 of which are commercially available to the public. A convenience sample of 200 emergency department (ED) patients with chief complaints of cardiopulmonary origin or a peripheral capillary oxygen saturation ≤ 94 percent were enrolled. Analysis of variance was performed to compare SpO2s and test characteristics of the 3 devices compared to control. The correlation between test devices and the control were high (r range 0.70-0.79). Although the test characteristics were not perfect, the devices did have good sensitivity using a cutoff value of 94% (sensitivity ranging from 90% to 92%), which improved with lower SpO2 cutoff values to 92% (sensitivity ranging from 96% to 97%). CONCLUSION: The 3 commercially available devices were accurate enough to be clinically useful when compared to a hospital bedside monitor pulse oximeter. Consumer-grade portable pulse oximeters may be useful if overwhelming numbers of patients require oxygen saturation monitoring, such as during the COVID-19 pandemic.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Schrading et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: