This article provides a methodological guide for using parallel-plate flow chambers to study thrombogenic potential in whole blood.
Answers are grounded in this paper and cite their sources.
Supports standardized in vitro thrombus assays; leaves open correlation with clinical thrombosis outcomes.
Custom-made and commercial parallel-plate flow chambers are widely used for studies of platelet activation and thrombus formation in whole blood at defined shear rates. When used in a reproducible way, such flow chamber devices give valuable information on the thrombogenic potential of human, mouse, or rat blood. This article aims to provide a practical guide for the use of parallel-plate flow chambers in combination with routine microscopic imaging techniques. The following methodological aspects are addressed: preparation of surface coatings, calculation of blood flow and shear rate, control of pre-analytical variables, protocols for routine performing of flow chamber tests with non-coagulating or coagulating blood, and procedures for real-time and end-point analysis of thrombus formation. Frequently encountered experimental problems and artifacts are discussed, as well as possibilities for using flow chamber devices as a diagnostic tool to test antithrombotic medication.
No takes yet. Share an insight, caveat, or question.
Kruchten et al. (2012) studied this question.