Key result
ICD lead dysfunction occurring >2.1 months post-LVAD is linked to ~367% higher persistent dysfunction risk.
Why the study?
ICD lead dysfunction after LVAD implantation can occur, but only short-term outcomes have been reported.
Cohort (n=170)
Yes
ICD lead dysfunction is highly prevalent (72.9%) in the long term after LVAD implantation, with late occurrence (>2.1 months) strongly predicting persistent dysfunction.
BACKGROUND: Implantable cardioverter defibrillator (ICD) lead dysfunction after left ventricular assist device (LVAD) implantation can occur, but only short-term outcomes have been reported. OBJECTIVES: We aimed to evaluate the long-term incidence, characteristics, and predictors of persistent ICD lead dysfunction after LVAD implantation. METHODS: This was a retrospective multicenter study. All patients with a transvenous ICD lead at the time of LVAD implantation between January 1, 2011, and December 31, 2023, were enrolled. The primary endpoint was lead dysfunction. Risk factors for persistent compared to temporary lead dysfunction using a logistic analysis were determined. RESULTS: A total of 170 patients (mean age: 48.0 ± 12.7 years) were analyzed. The median follow-up period was 46.2 (Q1-Q3: 32.3-61.4) months. Lead dysfunction was observed in 124 leads (72.9%), of which 60.4% (N = 75) occurred within a year after LVAD implantation. Of the 124 lead dysfunctions, 84 (67.7%) showed persistent dysfunction and 40 (32.3%) were temporary. Lead dysfunction occurring later than 2.1 months after LVAD implantation was a risk factor for persistent lead dysfunction (area under the curve: 0.673). Multivariable analysis identified lead dysfunction occurring later than 2.1 months after LVAD implantation as an independent risk factor for persistent lead dysfunction (OR: 4.67; 95% CI: 2.05-10.94; P < 0.001). CONCLUSIONS: In patients undergoing LVAD implantation, ICD lead dysfunction was observed in 72.9% after a mean follow-up of 46.2 months, of which two-thirds had persistence lead dysfunction. Later occurrence of lead dysfunction (>2 months after implant) may be a risk factor for persistent lead dysfunction.
No takes yet. Share an insight, caveat, or question.
Oshima et al. (2025) conducted a cohort in Advanced heart failure requiring Left Ventricular Assist Device (LVAD) implantation with pre-existing transvenous ICD (n=170). Left Ventricular Assist Device (LVAD) implantation vs. Temporary lead dysfunction was evaluated on Persistent lead dysfunction associated with occurrence >2.1 months post-LVAD implantation (OR 4.67, 95% CI 2.05-10.94, p=<0.001). In patients undergoing LVAD implantation, ICD lead dysfunction occurred in 72.9% of cases, and dysfunction occurring more than 2.1 months post-implant was an independent risk factor for persistent dysfunction (OR 4.67).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: