Key result
Continuous-flow LVAD implantation is linked to a ~21% drop in RV lead sensing amplitude.
Why the study?
Ventricular arrhythmias are common after continuous-flow LVAD implantation and may affect ICD performance, but changes in ICD parameters post-LVAD are not well characterized.
Does continuous-flow LVAD implantation affect the performance and adverse event rates of preexisting ICDs?
Cohort (n=44)
No
Does continuous-flow LVAD implantation affect the performance and adverse event rates of preexisting ICDs?
Continuous-flow LVAD implantation is associated with significant reductions in preexisting ICD right ventricular lead sensing amplitude and a high rate of ICD-related adverse events, warranting formal interrogation post-LVAD.
Answers are grounded in this paper and cite their sources.
LVAD implantation may reduce ICD sensing and prompt system modifications; leaves open whether routine post-implant interrogation improves outcomes.
BACKGROUND: Ventricular arrhythmias (VAs) in the months following continuous-flow left ventricular assist devices (LVADs) implantation are common and associated with increased morbidity and mortality. We sought to evaluate for changes in implantable cardioverter defibrillator (ICD) performance following LVAD implantation. METHODS: We retrospectively reviewed the ICD interrogation reports and medical records of patients who received a continuous-flow LVAD at the University of Chicago Medical Center between January 2008 and December 2010. Patients who did not have an ICD prior to LVAD implantation were excluded. RESULTS: Forty-four patients were included. Median right ventricular lead sensing amplitude decreased from 10.0 (interquartile range 7.6-13.6) mV prior to LVAD to 7.9 (5.9-9.3) mV after LVAD implantation (P < 0.01). Significant reductions in sensing threshold were associated with smaller lead diameters, smaller pace/sense electrode surface areas, and Medtronic ICD leads (Medtronic Inc., Minneapolis, MN, USA). Lead impedance and capture threshold did not change significantly. ICD-related adverse events occurred in 30% of patients and led to ICD modification in 18% of patients. CONCLUSIONS: Implantation of a continuous-flow LVAD is associated with changes in the performance of preexisting ICDs. ICD-related adverse events were encountered following LVAD implantation in this cohort, and at times resulted in invasive and noninvasive ICD system modification. Reductions in ICD sensing threshold after LVAD implantation may adversely affect ICD function. Formal ICD interrogation in addition to the regular follow-up testing is warranted post-LVAD.
No takes yet. Share an insight, caveat, or question.
Thomas et al. (2013) conducted a cohort in Preexisting ICD and continuous-flow LVAD (n=44). Continuous-flow LVAD implantation vs. Prior to LVAD implantation (baseline) was evaluated on Median right ventricular lead sensing amplitude (mV) (p=< 0.01). Continuous-flow LVAD implantation in patients with preexisting ICDs significantly decreased median right ventricular lead sensing amplitude from 10.0 mV to 7.9 mV (P<0.01).