Key result
A radiofrequency powered transseptal needle successfully achieved transseptal puncture in 35 of 36 attempts requiring RF energy, with transient right atrial thrombus occurring in 2 patients.
Why the study?
Does a radiofrequency powered transseptal needle safely and effectively facilitate transseptal puncture in patients undergoing left-sided catheter ablation?
Observational (n=35)
Does a radiofrequency powered transseptal needle safely and effectively facilitate transseptal puncture in patients undergoing left-sided catheter ablation?
An RF-powered transseptal needle provides a safe and effective method for achieving transseptal access during left-sided catheter ablation without requiring significant mechanical force.
May warrant caution with RF needles due to thrombus risk; leaves open need for controlled trials before routine use.
INTRODUCTION: The purpose of this study was to determine the safety and efficacy of using a novel radiofrequency (RF) powered transseptal needle to perform transseptal puncture (TSP). METHODS: TSP was performed in 35 consecutive patients undergoing left-sided catheter ablation (mean age = 51 years; male = 71%) using a RF powered transseptal needle (NRG, Adult Large and Standard Curve C1, 71 cm, Baylis Medical Company, Inc.). Prior TSP had been performed in 34% of patients. The transseptal apparatus was positioned with the tip of the dilator engaged in the fossa ovalis. RF energy was delivered to the tip of the transseptal needle using a proprietary RF generator at 10 W for 2 seconds as gentle pressure was applied to the needle. RESULTS: In 5 of the 41 TSPs, the needle crossed into the left atrium before RF energy was delivered. In 35 of the remaining 36 punctures, the needle was successfully advanced into the left atrium after application of RF current. In 1 patient, the TSP with the powered needle was unsuccessful but was accomplished using a standard needle. The only complication was a transient right atrial thrombus, which occurred in 2 patients. CONCLUSION: A radiofrequency powered transseptal needle can be used to perform TSP safely and successfully without the need for significant mechanical force, even in patients who have undergone TSP previously. Additional studies are needed to determine whether a powered transseptal needle should be used routinely.
No takes yet. Share an insight, caveat, or question.
Smelley et al. (2009) conducted an observational in Patients undergoing left-sided catheter ablation (n=35). Radiofrequency (RF) powered transseptal needle was evaluated on Successful transseptal puncture after application of RF current. A radiofrequency powered transseptal needle successfully achieved transseptal puncture in 35 of 36 attempts requiring RF energy, with transient right atrial thrombus occurring in 2 patients.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: