Key result
Anysis-200 shows high agreement with VerifyNow for assessing aspirin responsiveness, yielding ~92% sensitivity.
Why the study?
Analysis of responsiveness to antiplatelet therapy is crucial in cardiovascular disease management, prompting evaluation of the new Anysis-200 platelet function analyzer compared with VerifyNow.
Does the Anysis-200 system provide comparable accuracy to VerifyNow in assessing platelet aggregation and aspirin responsiveness in cardiology patients?
Observational (n=85)
Does the Anysis-200 system provide comparable accuracy to VerifyNow in assessing platelet aggregation and aspirin responsiveness in cardiology patients?
The Anysis-200 system demonstrates equivalent accuracy and significant agreement with VerifyNow for assessing aspirin responsiveness in cardiology patients.
Answers are grounded in this paper and cite their sources.
Anysis-200 may support aspirin responsiveness testing as an alternative to VerifyNow; leaves open prospective outcome validation before clinical adoption.
BACKGROUND: Analysis of responsiveness to antiplatelet therapy is crucial in the management of patients with cardiovascular diseases. OBJECTIVE: This study aimed to evaluate a new platelet function analysis system (Anysis-200) and to compare it with VerifyNow (Accumetrics, San Diego, CA, USA) in cardiology patients. METHODS: Overall, 125 citrated blood samples were collected from 85 cardiology patients referred for platelet function testing. In Anysis-200, platelet function was measured as blood migration distance (MD) until clogging of flow passage, which is comparable to aspirin resistance units obtained using VerifyNow. The two devices were simultaneously used and compared. RESULTS: The MDs before and after taking aspirin were 175±51 and 247±27 mm, respectively (p < 0.0001). Compared with VerifyNow (reference), the sensitivity and specificity of Anysis-200 was 91.5% and 75.5%, respectively (area under the curve, 0.829). Further, the true positive rate in patients newly taking aspirin was 85% for VerifyNow and 92.5% for Anysis-200, respectively. The Cohen's kappa coefficient between the two devices was 0.682, indicating a relatively high agreement. CONCLUSIONS: Anysis-200, a novel system for assessing platelet aggregation, has accuracy and precision equivalent to that of, and significant agreement with, VerifyNow. Anysis-200 may be useful in screening patients with abnormal platelet reactivity and aspirin nonresponsiveness.
No takes yet. Share an insight, caveat, or question.
Chung et al. (2020) conducted an observational in Cardiovascular disease (n=85). Anysis-200 vs. VerifyNow was evaluated on True positive rate in patients newly taking aspirin (AUC 0.829). The Anysis-200 platelet function analysis system showed high agreement with VerifyNow, with 91.5% sensitivity and 75.5% specificity for aspirin responsiveness.