Key result
Real time three-dimensional echocardiography strongly correlated with actual measurements for interventricular time difference and stroke volume in an in vitro porcine model.
Why the study?
Does RT3DE accurately measure interventricular dyssynchrony and stroke volume compared to actual measurements in an in vitro porcine model?
Does RT3DE accurately measure interventricular dyssynchrony and stroke volume compared to actual measurements in an in vitro porcine model?
RT3DE is a feasible and reproducible method to identify interventricular dyssynchrony and evaluate ventricular stroke volumes in an in vitro porcine model.
Answers are grounded in this paper and cite their sources.
RT3DE feasible for preclinical dyssynchrony models; leaves open clinical validation.
BACKGROUND: Loss of synchronous contraction between or within the right and left ventricle (RV, LV) leads to adverse ventricular function. We used real time three-dimensional echocardiography (RT3DE) for evaluation of severity of interventricular dyssynchrony and function in a porcine heart model. METHODS: Six fresh in vitro porcine hearts were used to create a controlled model of LV and RV dyssynchrony using two sets of pulsatile pumps. Synchronized and dyssynchronized pump settings were used with two different dyssynchronized settings: LV filled first and RV filled first. RESULTS: There was good correlation between actual measurement and RT3DE for interventricular time difference (r = 0.95, P < 0.0001) and stroke volume (SV) for LV and RV (0.89, 0.85; P < 0.0001, respectively). RT3DE data showed a small but significant underestimation for actual volume (P < 0.05). The intra- and interobserver variabilities are 2.9 +/- 1.5% and 3.1 +/- 5.4% for LV and RV SVs, and 1.7 +/- 2.4% and 2.2 +/- 3.2% for time differences by RT3DE. There was significant difference in RV SV between synchrony and dyssynchrony when the RV filled first (P < 0.05), but not in other groups. The same pattern was found in RT3DE derived SVs (synchrony versus dyssynchrony with RV filled first, P < 0.05). CONCLUSIONS: There is no compromise in LV SV during interventricular dyssynchrony; but RV SV was significantly diminished when the RV filled first. RT3DE is a feasible, robust and reproducible method to identify interventricular dyssynchrony and to evaluate ventricular SVs.
No takes yet. Share an insight, caveat, or question.
Zheng et al. (2010) studied Interventricular dyssynchrony (n=6). Real time three-dimensional echocardiography (RT3DE) vs. Actual measurement was evaluated on Correlation between actual measurement and RT3DE for interventricular time difference (r = 0.95, p=< 0.0001). Real time three-dimensional echocardiography strongly correlated with actual measurements for interventricular time difference and stroke volume in an in vitro porcine model.
Synapse has enriched 10 related papers. Consider them for comparative context: