Key result
Electrocardiographic imaging feasibly measures novel repolarization parameters, revealing steep gradients in regions with flat electrograms.
Why the study?
Measuring repolarization characteristics is challenging, and the potential of novel repolarization parameters in ECGI for arrhythmic risk stratification warranted exploration.
Does the use of novel repolarization parameters in electrocardiographic imaging uncover arrhythmogenic substrate in IVF patients?
Observational (n=8)
Does the use of novel repolarization parameters in electrocardiographic imaging uncover arrhythmogenic substrate in IVF patients?
Measuring novel repolarization parameters using ECGI is feasible, fully automated, and may help identify arrhythmogenic substrates for risk stratification.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for repolarization imaging in IVF substrate detection; larger prospective studies needed before clinical adoption.
BACKGROUND: Measuring repolarization characteristics is challenging and has been reserved for experienced physicians. In electrocardiographic imaging (ECGI), activation-recovery interval (ARI) is used as a measure of local cardiac repolarization duration. We hypothesized that repolarization characteristics, such as local electrogram morphology and local and global dispersion of repolarization timing and duration could be of significance in ECGI. OBJECTIVE: To further explore their potential in arrhythmic risk stratification we investigated the use of novel repolarization parameters in ECGI. MATERIALS AND METHODS: We developed and compared methods for T-peak and T-end detection in reconstructed potentials. All methods were validated on annotated reconstructed electrograms (EGMs). Characteristics of the reconstructed EGMs and epicardial substrate maps in IVF patients were analyzed by using data recorded during sinus rhythm. The ECGI data were analyzed for EGM morphology, conduction, and repolarization. RESULTS: We acquired ECGI data from 8 subjects for this study. In all patients we evaluated four repolarization parameters: Repolarization time, T-wave area, Tpeak-Tend interval, and T-wave alternans. Most prominent findings were steep repolarization time gradients in regions with flat EGMs. These regions were also characterized by low T-wave area and large differences in Tpeak-Tend interval. CONCLUSIONS: Measuring novel repolarization parameters in reconstructed electrograms acquired with ECGI is feasible, can be done in a fully automated manner and may provide additional information on underlying arrhythmogenic substrate for risk stratification. Further studies are needed to investigate their potential use and clinical application.
No takes yet. Share an insight, caveat, or question.
Blom et al. (2020) conducted an observational in Idiopathic ventricular fibrillation (IVF) (n=8). Electrocardiographic imaging (ECGI) repolarization parameters was evaluated on Feasibility of measuring novel repolarization parameters (Repolarization time, T-wave area, Tpeak-Tend interval, and T-wave alternans). Measuring novel repolarization parameters using electrocardiographic imaging was feasible in 8 subjects, revealing steep repolarization time gradients in regions with flat electrograms.
Synapse has enriched 10 related papers. Consider them for comparative context: