Key result
Attain Stability active fixation coronary sinus leads linked to ~34% fewer total displacements versus passive leads.
Why the study?
A novel active fixation coronary sinus lead was released to improve targeted lead positioning, but its implant procedure parameters and electromechanical stability needed comparison with passive leads.
Does an active fixation coronary sinus lead reduce total lead displacements compared to passive fixation leads in patients receiving coronary sinus leads?
Cohort (n=761)
Yes
Does an active fixation coronary sinus lead reduce total lead displacements compared to passive fixation leads in patients receiving coronary sinus leads?
Active fixation coronary sinus leads improve electromechanical stability and reduce lead displacements compared to passive leads, with minimal increases in procedure and fluoroscopy times.
Answers are grounded in this paper and cite their sources.
Active fixation CS leads were associated with fewer displacements; leaves open whether they should be preferred pending randomized confirmation.
Background A novel active fixation coronary sinus (CS) lead, Attain Stability (AS), has been released aiming to improve targeted lead positioning. Rather than being wedged into the distal vessel, it relies on a side helix for fixation. We aimed to compare implant procedure parameters and electromechanical stability of the AS lead with passive CS leads. Methods A retrospective study involving six major UK cardiac centers. Patients who received active fixation leads were compared with passive fixation lead recipients in a 1:2 ratio. The primary outcome was total lead displacements (combined macrodisplacement/microdisplacement, defined as displacements requiring repositioning procedures, an increase in threshold ≥0.5 V or pulse width ≥0.5 ms, or a change in pacing polarity). Results A total of 761 patients were included (253 AS leads and 508 passive fixation leads), of which 736 had follow‐up data. The primary endpoint rate was 31% (75/241) in the active and 43% (214/495) in the passive group ( p = .002). Six patients (2.5%) in the active group and 14 patients (2.8%) in the passive group required CS lead repositioning procedures ( p = 0.981). On multivariable analysis, active leads were associated with a reduction in lead displacements, odds ratio 0.66 (95% confidence interval: 0.46–0.95), p = .024. There were differences in favor of passive leads in procedure duration, 120 (96–149) versus 127 (105–155) min ( p = .008), and fluoroscopy time, 17 (11–26) versus 18.5 (13–27) min ( p = .0022). The median follow‐up duration was similar (active vs. passive): 31 (17–47) versus 34 (16–71) weeks, ( p = .052). Conclusion AS CS leads had improved electromechanical stability compared with passive fixation leads, with only minimal increases in implant procedure and fluoroscopy times.
No takes yet. Share an insight, caveat, or question.
Gerontitis et al. (2020) conducted a cohort in Patients receiving coronary sinus leads (n=761). Attain Stability active fixation coronary sinus lead vs. Conventional passive fixation coronary sinus leads was evaluated on Total lead displacements (combined macrodisplacement/microdisplacement) (OR 0.66, 95% CI 0.46-0.95, p=0.024). The Attain Stability active fixation coronary sinus lead reduced total lead displacements compared with passive fixation leads (31% vs 43%; OR 0.66; 95% CI 0.46-0.95; p=0.024).