Key result
Extracting the Medtronic active fixation LV lead at 8 months appears safe without coronary sinus damage.
Why the study?
Is the transvenous extraction of the novel active fixation coronary sinus lead safe and effective?
Case Report (n=1)
No
Is the transvenous extraction of the novel active fixation coronary sinus lead safe and effective?
The transvenous extraction of the novel Medtronic 20066 Attain Stability active fixation LV lead appears safe and effective at 8 months post-implantation.
Hypothesis-generating for extraction of this novel active fixation CS lead; does not yet support routine clinical adoption.
AIMS: Benefits of cardiac resynchronization therapy (CRT) are well known for heart failure; however, some patients might experience complications related to the coronary sinus (CS) lead (high pacing threshold, phrenic nerve stimulation, and dislodgment) with unfavourable impact on quality of life, costs, and management. Lead stability is one of the most common unmet needs for CRT procedures. METHODS AND RESULTS: Recently, new model Medtronic 20066 Attain Stability(®) (Maastricht, The Netherlands) active fixation LV lead has been released, to overcome this issue. The lead has a small side helix of 0.20 mm (0.008 in.) that allows for secure placement of the lead within the vein at the desired location. We report our first experience with the extraction of this novel active fixation left ventricular lead. CONCLUSION: In our case, to our knowledge the first reported in humans, the extraction of this new model of active fixation lead was proved to be a safe and effective procedure at 8 months after implantation. Indeed, under angiographic and fluoroscopic check, there was no documented dissection or damage to the CS during and after removal of the lead. The rotation manoeuvre was effective when combined with moderate traction of the lead itself.
No takes yet. Share an insight, caveat, or question.
Bontempi et al. (2015) conducted a case report in Heart failure requiring cardiac resynchronization therapy (n=1). Transvenous extraction of active fixation left ventricular lead was evaluated on Safety and efficacy of the extraction procedure. Extraction of the novel Medtronic 20066 Attain Stability active fixation left ventricular lead at 8 months post-implantation was safe and effective, with no documented damage to the coronary sinus.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: