Key result
The Attain StarFix 4195 active fixation coronary sinus lead demonstrates improved dislodgment rates but lower extractability rates, even after a short dwelling period.
Case Report
While active fixation coronary sinus leads improve stability and reduce dislodgment in CRT, they pose significant challenges for transvenous extraction even after short implant durations.
Answers are grounded in this paper and cite their sources.
Extraction of active fixation CS leads may fail even shortly after implantation; leaves open questions on long-term extractability and patient selection.
Cardiac resynchronization therapy (CRT) has been shown to reduce morbidity and mortality in select patients with heart failure with reduced left ventricle ejection fraction (LVEF); however, in about 10% of patients, CRT fails owing to coronary sinus (CS) lead dislodgment.1–4 Over the past 15 years, active fixation CS leads have been developed to address the need for greater stability of the CS lead.3 Attain StarFix 4195 (Medtronic, Minneapolis, MN), a 5F unipolar steroid-eluting CS lead with an extendable lobe for lead active fixation, demonstrated a significant improvement in dislodgment rates, but with a lower extractability rate, even after a short dwelling period.
No takes yet. Share an insight, caveat, or question.
Cori et al. (2024) conducted a case report in Heart failure with reduced left ventricle ejection fraction (LVEF). Attain StarFix 4195 active fixation coronary sinus lead was evaluated. The Attain StarFix 4195 active fixation coronary sinus lead demonstrates improved dislodgment rates but lower extractability rates, even after a short dwelling period.