Key result
The MESI ABPI MD device successfully passed the ESH-IP validation protocol for hypertension screening with a systolic blood pressure bias of -3.5 mmHg, whereas the Dopplex ABIlity failed.
Why the study?
It was unclear whether brachial blood pressure measurements from automated PAD screening systems are sufficiently accurate for simultaneous hypertension screening in cardiovascular screening programmes.
Are automated ankle-brachial pressure index (ABPI) measurement devices accurate for simultaneous hypertension screening compared to a reference sphygmomanometer?
Cross-Sectional (n=33)
No
Are automated ankle-brachial pressure index (ABPI) measurement devices accurate for simultaneous hypertension screening compared to a reference sphygmomanometer?
The MESI ABPI MD automated device is sufficiently accurate for simultaneous hypertension and peripheral artery disease screening, potentially streamlining population cardiovascular screening programs.
Answers are grounded in this paper and cite their sources.
Supports MESI ABPI MD accuracy for combined screening; leaves open prospective trials before routine use.
OBJECTIVE: In a large-scale population cardiovascular screening programme, peripheral artery disease (PAD) and hypertension would ideally be rapidly assessed using a single device. The ankle-brachial pressure index (ABPI) is calculated by comparing the ankle and brachial blood pressure (BP). However, it is currently unclear whether brachial BP measurements provided by automated PAD screening systems are sufficiently accurate for simultaneous hypertension screening. METHODS: Two portable PAD screening devices, the MESI ABPI MD and Huntleigh's Dopplex ABIlity, were evaluated following the European Society of Hypertension International Protocol (ESH-IP) Revision 2010 using a mercury-free sphygmomanometer as a reference device. RESULTS: On average, the MESI slightly underestimated brachial systolic blood pressure (BP) with a bias and standard deviation (SD) of -3.5 (SD: 3.3) mmHg and diastolic BP with a bias of -1.5 (SD: 2.3) mmHg. For systolic BP estimates, the Dopplex was more accurate than the MESI with a lower bias of -0.5 (SD: 4.2) mmHg but less precise. The MESI successfully fulfilled all the requirements of the ESH-IP for hypertension screening. The Dopplex device failed the ESH-IP due to the absence of DBP measurements. CONCLUSIONS: The MESI device appears to be suitable for simultaneous PAD and hypertension screening as part of a preventative care programme. Huntleigh's Dopplex ABIlity failed to pass the ESH-IP validation test. Further clinical trials are underway to assess the use of the MESI for simultaneous screening for hypertension and PAD in a population screening setting.
No takes yet. Share an insight, caveat, or question.
Janus et al. (2023) conducted a cross-sectional in Hypertension screening (n=33). Automated PAD screening devices (MESI ABPI MD and Dopplex ABIlity) vs. Nissei DM-3000 sphygmomanometer was evaluated on Mean bias in systolic blood pressure (MESI vs reference) (Mean bias -3.5 mmHg, 95% CI -10.0 to 3.0). The MESI ABPI MD device successfully passed the ESH-IP validation protocol for hypertension screening with a systolic blood pressure bias of -3.5 mmHg, whereas the Dopplex ABIlity failed.
Synapse has enriched 10 related papers. Consider them for comparative context: