Key result
3D echocardiography aligns better with thoracic CT than fluoroscopy for evaluating RV non-apical lead position.
Why the study?
RV septal pacing is considered better than traditional apical pacing, but agreement among imaging modalities for evaluating the non-apical lead tip needed investigation.
Does three-dimensional echocardiography agree with computed tomography and fluoroscopy in evaluating the tip of the RV pacing lead in the non-apical position in patients with permanent pacemaker implantation?
Observational (n=54)
Does three-dimensional echocardiography agree with computed tomography and fluoroscopy in evaluating the tip of the RV pacing lead in the non-apical position in patients with permanent pacemaker implantation?
Effect estimate: Kappa 0.734
Three-dimensional echocardiography shows high agreement with thoracic CT for evaluating RV septal lead position, suggesting it is a useful tool for defining and guiding RV non-apical lead positioning.
May support 3D echo for RV non-apical lead assessment; leaves open outcome validation versus fluoroscopy.
Objective Right ventricular (RV) septal pacing is considered a better pacing procedure compared with traditional apical pacing. This study aimed to investigate agreement among computed tomography (CT), three-dimensional echocardiography (3D-echo), and fluoroscopy for evaluating the tip of the RV pacing lead in the non-apical position in patients with permanent pacemaker implantation. Methods Fifty-four patients were prospectively enrolled. Data on patients’ characteristics and imaging findings were analyzed. The agreement rate in distinguishing the RV septal lead position among the three imaging modalities was determined. Results Thirty-three (61%) patients were men and the median age was 76 years. Overall, the agreement rate among the three imaging modalities was 87% (47/54; Kappa ratio: 0.734). The agreement of 3D-echo compared with thoracic CT (Kappa ratio: 0.893) was better than that for thoracic CT and fluoroscopy (Kappa ratio: 0.658). Agreement between fluoroscopy and 3D-echo was lowest (Kappa ratio: 0.632). Conclusions Agreement in evaluating the position of the septal lead between thoracic CT and 3D-echo is better than that between other imaging modalities. Our findings indicate that 3D-echo imaging might be the best imaging tool for defining the tip of the RV non-apical lead position and be useful for guiding positioning of the RV lead.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2021) conducted an observational in Permanent pacemaker implantation (n=54). Three-dimensional echocardiography (3D-echo) vs. Computed tomography (CT) and fluoroscopy was evaluated on Agreement rate in distinguishing the RV septal lead position among the three imaging modalities (Kappa 0.734). Three-dimensional echocardiography showed better agreement with thoracic CT (Kappa 0.893) than fluoroscopy for evaluating right ventricular non-apical lead position.