Key result
3D transthoracic echocardiography achieves ~80% agreement with CT for right ventricular lead tip localization.
Why the study?
Right ventricular leads are routinely assessed with fluoroscopy or chest radiography after pacemaker or defibrillator implantation, and the authors hypothesized that 3D transthoracic echocardiography would offer better lead tip localization comparable to computed tomography.
Does 3D transthoracic echocardiography improve right ventricular lead tip localization compared to fluoroscopy and chest radiography, and perform comparably to computed tomography, in patients with pacemaker or defibrillator leads?
Observational (n=50)
Does 3D transthoracic echocardiography improve right ventricular lead tip localization compared to fluoroscopy and chest radiography, and perform comparably to computed tomography, in patients with pacemaker or defibrillator leads?
Answers are grounded in this paper and cite their sources.
3D transthoracic echocardiography performs comparably to CT and better than fluoroscopy or chest radiography for localizing RV lead tips in the short-axis, offering a potential tool for real-time visualization during lead delivery.
Right ventricular (RV) leads are targeted to the septum during pacemaker or defibrillator implantation to reduce risk of perforation and limit dyssynchrony. While routinely assessed with fluoroscopy or chest radiography (CXR), we hypothesized that 3D transthoracic echocardiography (3DE) would offer better lead tip localization, comparable to computed tomography (CT). CT chest, full-volume 3DE, fluoroscopy, and post-implant CXRs were independently assessed. Lead tip location was characterized using a 12-segment model (short-axis: septal, inferior, lateral, and anterior; long-axis: base, mid, or apex). Kappa coefficients (κ) were calculated to assess agreement. CXR, CT and 3DE were analyzed in 50 patients. Apical long-axis location was identified in 92% by CXR, and 96% by 3DE or CT. Septal short-axis location was found in 20% by CXR, 56% by 3DE, and 62% by CT. Overall 3-modality agreement in the long-axis was good (κ = 0.61, p < 0.001), and fair for the short-axis (κ = 0.28, p < 0.001). 3DE and CT demonstrated 80% pair-wise agreement (κ = 0.67, p < 0.001). Among 32 patients with all 4 modalities, septal short-axis location was identified in 81% by fluoroscopy, 25% by CXR, 63% by CT scan, and 53% by 3DE. Overall, inter-modality agreement for long-axis was moderate (κ = 0.48, p < 0.001) and in the short-axis was poor (κ = 0.16, p < 0.001). Localizing the RV lead tip remains challenging, particularly in the short-axis, where imaging with fluoroscopy or CXR may be unreliable. 3DE performed similarly to CT in lead tip localization, and may present an opportunity for real-time visualization of lead delivery.
No takes yet. Share an insight, caveat, or question.
Addetia et al. (2026) conducted an observational in Pacemaker or defibrillator implantation with right ventricular lead (n=50). 3D transthoracic echocardiography (3DE) vs. Computed tomography (CT), fluoroscopy, and chest radiography (CXR) was evaluated on Right ventricular lead tip localization and inter-modality agreement (12-segment model) (κ = 0.67 (3DE vs CT pair-wise agreement: 80%), p=<0.001). 3D transthoracic echocardiography demonstrated 80% pair-wise agreement with CT in localizing right ventricular lead tips, identifying septal position in by CT.
Synapse has enriched 10 related papers. Consider them for comparative context: