Key result
CT, TTE, and CXR identified RV lead tip position in 100% of patients, whereas MRI failed in 39% due to artifact, with CT providing the clearest definition of lead tip position.
Why the study?
Does CT, MRI, TTE, or CXR provide the optimal assessment of right ventricular pacing lead position in patients with MRI-conditional pacemakers?
Cross-Sectional (n=18)
Does CT, MRI, TTE, or CXR provide the optimal assessment of right ventricular pacing lead position in patients with MRI-conditional pacemakers?
Absolute Event Rate: 100% vs 61%
CT provides the clearest definition of right ventricular pacing lead position compared to MRI, TTE, and CXR, revealing that many leads intended for the septum are not truly septal.
BACKGROUND: Right ventricular nonapical (RVNA) pacing may reduce the risk of heart failure. Fluoroscopy is the standard approach to determine lead tip position, but is inaccurate. We compared cardiac computed tomography (CT), magnetic resonance imaging (MRI), two-dimensional and three-dimensional transthoracic echocardiography (TTE), and chest x-ray (CXR) to assess which provides the optimal assessment of right ventricular (RV) lead tip position. METHODS: Eighteen patients with MRI-conditional pacemakers (10 RVNA and eight apical [RVA] leads) underwent contrast CT, MRI, TTE, and a standard postimplant posteroanterior and lateral CXR. To compare images, the RV was arbitrarily partitioned into three long-axis segments (right ventricular outflow tract, middle, and apex), and two short-axis segments (septal and nonseptal). Agreement between modalities was assessed. RESULTS: RV lead tip position was identified in all patients on CT, TTE, and CXR, but was not identified in seven (39%) patients on MRI due to device-related artifact. Of 10 leads deemed to be nonapical/septal during implant, 70% were identified as nonapical on CXR, 60% on CT, 60% on MRI, and 80% on TTE. On CT imaging only 10% were truly septal, 20% on MRI, 30% on CXR, and 80% on TTE. Agreement was better between modalities when assessing position of the designated RVA leads. CONCLUSION: During implant leads intended for the septum are not confirmed as such on subsequent imaging, and marked heterogeneity is apparent between modalities. MRI is limited by artifact, and discrepancy exists between TTE and CT in identifying septal lead position. CT gave the clearest definition of lead tip position.
No takes yet. Share an insight, caveat, or question.
Moore et al. (2016) conducted a cross-sectional in MRI-conditional pacemakers (n=18). Cardiac computed tomography (CT) vs. MRI, TTE, and CXR was evaluated on Identification of right ventricular lead tip position. CT, TTE, and CXR identified RV lead tip position in 100% of patients, whereas MRI failed in 39% due to artifact, with CT providing the clearest definition of lead tip position.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: