Key result
Perpendicular electrode placement and longer helices optimize HBP electrical resynchronization in simulation and ex-vivo models.
Why the study?
His bundle pacing is an alternative for cardiac resynchronization therapy, but the optimal electrode placement and pacing protocol for complete His-Purkinje block and LBBB required assessment.
Does varying electrode placement and pacing protocol optimize His bundle pacing for electrical synchrony in LBBB models?
Does varying electrode placement and pacing protocol optimize His bundle pacing for electrical synchrony in LBBB models?
In a computational and ex-vivo model of LBBB, perpendicular electrode placement and increased helix length optimized electrical synchrony during His bundle pacing.
Answers are grounded in this paper and cite their sources.
May inform lead design for His bundle pacing; leaves open prospective human validation.
His bundle pacing (HBP) has emerged as a feasible alternative to right (RVP) and biventricular pacing (BVP) for Cardiac Resynchronization Therapy (CRT). This study sought to assess, in ex-vivo experimental models, the optimal setup for HBP in terms of electrode placement and pacing protocol to achieve superior electrical synchrony in the case of complete His-Purkinje block and left bundle branch block (LBBB). We developed a 3D model of His bundle and bundle branches, embedded in a patient-specific biventricular heart model reconstructed from CT images. A monodomain reaction-diffusion model was adopted to describe the propagation of cardiac action potential, and a custom procedure was developed to compute pseudo-ECGs. Experimental measurements of tip electrode potential waveforms have been performed on ex-vivo swine myocardium to determine the appropriate boundary condition for delivering the electrical stimulus in the numerical model. An extended parametric analysis, investigating the effect of the electrode orientation and helix length, pacing protocol, and atrioventricular delay, allowed us to determine the optimal setup for HBP therapy. Both selective (S-HBP) and non-selective (NS-HBP) His bundle pacing were tested, as the variable anatomical location of the His bundle can result in the activation of the surrounding myocardium. Our study indicates a perpendicular placement of the electrode as the most advantageous for restoring the physiological function of the His-Purkinje system. We found that higher-energy protocols can compensate for the effects of an angled placement though concurring to potential tip fibrosis. Promisingly, we also revealed that an increased electrode helix length can provide optimal resynchronization even with low-energy pacing protocols. Our results provide informative guidance for implant procedure and therapy optimization, which will hopefully have clinical implications further improving the procedural success rates and patients' quality of life, due to reduced incidence of lead revision and onset of complications.
No takes yet. Share an insight, caveat, or question.
Barone et al. (2022) studied Complete His-Purkinje block and left bundle branch block (LBBB). His bundle pacing (HBP) was evaluated on Optimal setup for HBP to achieve superior electrical synchrony. Perpendicular electrode placement and increased helix length provided optimal electrical resynchronization for His bundle pacing in 3D simulation and ex-vivo models.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: