April 15, 1995Circulation627 citations

Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle Preparation

View Full Paper
HNHiroshi NakagawaWYWilliam S. YamanashiJPJan Piťha

Key Result

Saline-irrigated radiofrequency ablation produced deeper lesions (9.9 mm) than temperature-controlled (6.1 mm) and constant voltage (4.7 mm) ablation without irrigation (P<0.01).

Key Points

  • This study aims to determine if saline irrigation during radiofrequency ablation increases lesion depth and alters tissue temperature profiles.
  • Conducted in 11 anesthetized dogs with thigh muscle exposure.
  • Delivered radiofrequency current at various settings across three groups: constant voltage without irrigation, temperature control without irrigation, and constant voltage with saline irrigation.
  • Measured electrode and tissue temperatures at varying depths during procedures.
  • In the saline irrigation group, mean lesion depth was 9.9 mm with a volume of 700 mm3, significantly larger than 4.7 mm and 135 mm3 in the CV group (P < 0.01).
  • Tissue temperature at 3.5 mm depth in the CV irrigation group averaged 94.7 degrees C compared to lower temperatures in control groups (62.1 and 67.9 degrees C).
  • Saline irrigation effectively prevented impedance rise while allowing higher power delivery, resulting in larger lesions.

Structured PICO

Does saline-irrigated radiofrequency ablation increase lesion depth and volume compared to non-irrigated or temperature-controlled ablation in a canine thigh muscle model?

P
Population
11 anesthetized dogs, thigh muscle exposed and bathed with heparinized canine blood (36-37 degrees C)
I
Intervention
Radiofrequency ablation at high constant voltage (66 V) with saline irrigation through the catheter lumen and ablation electrode at 20 mL/min for up to 60 seconds
C
Comparator
Radiofrequency ablation at high constant voltage (66 V) without irrigation, and variable voltage (20 to 66 V) to maintain tip-electrode temperature at 80-90 degrees C without irrigation
O
Outcome
Tissue temperature profile (at depths of 3.5 and 7.0 mm) and lesion geometry (depth, maximal diameter, volume)surrogate

Saline-irrigated radiofrequency ablation allows for higher power delivery by preventing impedance rises, resulting in deeper and larger lesions due to deeper direct resistive heating.

Main Result

Absolute Event Rate: 9.9% vs 6.1%

p-value: p=<.01

Abstract

BACKGROUND: It is thought that only a thin layer of tissue adjacent to the electrode is heated directly by electrical current (resistive heating) during radiofrequency ablation. Most of the thermal injury is thought to result from conduction of heat from the surface layer. The purpose of this study was to determine whether lesion depth could be increased by producing direct resistive heating deeper in the tissue with higher radiofrequency power, allowed by cooling the ablation electrode with saline irrigation to prevent the rise in impedance that occurs when the electrode-tissue interface temperature reaches 100 degrees C. METHODS AND RESULTS: In 11 anesthetized dogs, the thigh muscle was exposed and bathed with heparinized canine blood (36 degrees C to 37 degrees C). A 7F catheter, with a central lumen, a 5-mm tip electrode with six irrigation holes, and an internal thermistor, was positioned perpendicular to the thigh muscle and held at a constant contact weight of 10 g. Radiofrequency current was delivered to 145 sites (1) at high constant voltage (66 V) without irrigation (CV group, n = 31), (2) at variable voltage (20 to 66 V) to maintain tip-electrode temperature at 80 degrees C to 90 degrees C without irrigation (temperature-control group, n = 39), and (3) at high CV (66 V) with saline irrigation through the catheter lumen and ablation electrode at 20 mL/min (CV irrigation group, n = 75). Radiofrequency current was applied for 60 seconds but was terminated immediately in the event of an impedance rise > or = 10 omega. Tip-electrode temperature and tissue temperature at depths of 3.5 and 7.0 mm were measured in all three groups (n = 145). In 33 CV irrigation group applications, temperature was also measured with a separate probe at the center (n = 18) or edge (n = 15) of the electrode-tissue interface. In all 31 CV group applications, radiofrequency energy delivery was terminated prematurely (at 11.6 +/- 4.8 seconds) owing to an impedance rise associated with an electrode temperature of 98.8 +/- 2.1 degrees C. All 39 temperature-control applications were delivered for 60 seconds without an impedance rise, but voltage had to be reduced to 38.4 +/- 6.1 V to avoid temperatures > 90 degrees C (mean tip-electrode temperature, 84.5 +/- 1.4 degrees C). In CV irrigation applications, the tip-electrode temperature was not > 48 degrees C (mean, 38.4 +/- 5.1 degrees C) and the electrode-tissue interface temperature was not > 80 degrees C (mean, 69.4 +/- 5.7 degrees C). An abrupt impedance rise with an audible pop and without coagulum occurred in 6 of 75 CV irrigation group applications at 30 to 51 seconds, probably owing to release of steam from below the surface. In the CV and temperature-control group applications, the temperatures at depths of 3.5 (62.1 +/- 15.1 degrees C and 67.9 +/- 7.5 degrees C) and 7.0 mm (40.3 +/- 5.3 degrees C and 48.3 +/- 4.8 degrees C) were always lower than the electrode temperature. Conversely, in CV irrigation group applications, electrode and electrode-tissue interface temperatures were consistently exceeded by the tissue temperature at depths of 3.5 mm (94.7 +/- 9.1 degrees C) and occasionally 7.0 mm (65.1 +/- 9.7 degrees C). Lesion dimensions were smallest in CV group applications (depth, 4.7 +/- 0.6 mm; maximal diameter, 9.8 +/- 0.8 mm; volume, 135 +/- 33 mm3), intermediate in temperature-control group applications (depth, 6.1 +/- 0.5 mm; maximal diameter, 11.3 +/- 0.9 mm; volume, 275 +/- 55 mm3), and largest in CV irrigation group applications (depth, 9.9 +/- 1.1 mm; maximal diameter, 14.3 +/- 1.5 mm; volume, 700 +/- 217 mm3; P < .01, respectively). CONCLUSIONS: Saline irrigation maintains a low electrode-tissue interface temperature during radiofrequency application at high power, which prevents an impedance rise and produces deeper and larger lesions. A higher temperature in the tissue (3.5 mm deep) than at the electrode-tissue interface indicates that direct resistive heating occurred deeper

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Nakagawa et al. (1995) studied Radiofrequency ablation (n=11). Radiofrequency ablation with saline irrigation vs. Temperature control and constant voltage without irrigation was evaluated on Lesion depth (mm) (p=<.01). Saline-irrigated radiofrequency ablation produced deeper lesions (9.9 mm) than temperature-controlled (6.1 mm) and constant voltage (4.7 mm) ablation without irrigation (P<0.01).