Key result
Radiofrequency ablation using a 2-mm irrigated tip electrode produced deeper lesions (8.0 vs 5.4 mm) and higher tissue temperatures than a 5-mm electrode in perpendicular orientation.
Why the study?
Does a 2-mm irrigated tip electrode increase lesion size and tissue temperature compared to a 5-mm electrode during radiofrequency ablation?
Population
11 anesthetized dogs (thigh muscle preparation bathed with heparinized canine blood, 148 ablation sites)
Comparison
Radiofrequency ablation using a 7F catheter with… vs Radiofrequency ablation using a 7F catheter with…
Design
Preclinical
Authors
Loading...
May guide irrigated catheter design; leaves open translation to human RF ablation outcomes.
Does a 2-mm irrigated tip electrode increase lesion size and tissue temperature compared to a 5-mm electrode during radiofrequency ablation?
Absolute Event Rate: 8% vs 5.4%
During radiofrequency ablation with active cooling, a smaller 2-mm electrode creates larger, deeper lesions with less dependence on orientation compared to a 5-mm electrode.
Nakagawa et al. (1998) studied Radiofrequency ablation (n=11). 2-mm irrigated tip electrode vs. 5-mm irrigated tip electrode was evaluated on Lesion depth in perpendicular orientation (mm). Radiofrequency ablation using a 2-mm irrigated tip electrode produced deeper lesions (8.0 vs 5.4 mm) and higher tissue temperatures than a 5-mm electrode in perpendicular orientation.