Repetitive scratching triggers Twiddler's syndrome in an ICD patient treated by lead replacement and fixation.
Case Report (n=1)
Highlights behavioral factors such as scratching and abnormal movements at the implantation site as a trigger for Twiddler's syndrome in ICD recipients, managed via lead replacement and fascial fixation.
Answers are grounded in this paper and cite their sources.
Abnormal scratching may trigger Twiddler's syndrome in ICD patients; this case leaves open the prevalence of behavioral causes.
A 72-year-old male with severe dilated cardiomyopathy, benefited of implantable cardiac defibrillator implementation. Device control shows high impedance. On X-ray, electrodes were completely twisted in the generator pocket, they were replaced and the generator was fixed to pectoralis-major fascia. Nurses report patient abnormal movements, scratching implantation area. This was identified as probably the trigger of the complication.
No takes yet. Share an insight, caveat, or question.
Minaoui et al. (2022) conducted a case report in Severe dilated cardiomyopathy and Twiddler's syndrome (n=1). Electrode replacement and generator fixation to pectoralis-major fascia following ICD implantation was evaluated on Resolution of Twiddler's syndrome (twisted electrodes and high device impedance). Abnormal patient movements and scratching at the ICD implantation site triggered Twiddler's syndrome (n=1), requiring electrode replacement and generator fixation.