Key result
A 69-year-old female developed twiddler's syndrome four months after CRT-D implantation, with the three leads meticulously braided in front of the sutured can.
Case Report (n=1)
This case report highlights a classic presentation of twiddler's syndrome, a permanent device malfunction caused by patient manipulation, in a patient with a CRT-D.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to consider manipulation syndromes in CRT-D malfunction; extends case literature while leaving prevention strategies open.
A 69-year-old obese and depressive female patient with symptomatic dilated cardiomyopathy was implanted with cardiac resynchronization therapy in combination with an implantable cardioverter-defibrillator in primary prevention. Four months later, the patient was referred to our centre for suspected lead dislodgment. A typical twiddler's syndrome became obvious when we opened the pocket: the can was still sutured on to the muscle while the three leads were nicely and meticulously braided just in front of it. First described in 1968, twiddler's syndrome refers to permanent malfunction of the device due to the patient's manipulation. Conflict of Interest: none declared.1
No takes yet. Share an insight, caveat, or question.
Bovéda et al. (2012) conducted a case report in Symptomatic dilated cardiomyopathy (n=1). Cardiac resynchronization therapy with an implantable cardioverter-defibrillator was evaluated on Lead dislodgment / twiddler's syndrome. A 69-year-old female developed twiddler's syndrome four months after CRT-D implantation, with the three leads meticulously braided in front of the sutured can.