Key result
The CARTO system yielded comparable left ventricular volumes to CMR but larger right ventricular volumes, questioning their clinical interchangeability.
Why the study?
Does volume measurement by the CARTO electrophysiological mapping system provide interchangeable results with cardiac magnetic resonance for ventricular volumes in heart failure patients?
Cross-Sectional (n=34)
Does volume measurement by the CARTO electrophysiological mapping system provide interchangeable results with cardiac magnetic resonance for ventricular volumes in heart failure patients?
While CARTO and CMR show good correlation for ventricular volumes, they are not clinically interchangeable, particularly for right ventricular measurements.
Answers are grounded in this paper and cite their sources.
CARTO and CMR ventricular volumes may not be interchangeable in HF; leaves open need for outcome-based validation.
AIMS: The CARTO electrophysiological mapping system has demonstrated accurate results for end-diastolic ventricular volumes in casts and animals. However, in humans, a comparison with cardiac magnetic resonance (CMR), the non-invasive gold standard for volumetric analysis, has not yet been performed. METHODS AND RESULTS: A total of 34 (29 male) heart failure patients (NYHA class III/IV) underwent an electrophysiological mapping procedure with the CARTO system in the left ventricle (LV) (n = 34) and right ventricle (RV) (n = 12) and CMR for RV and LV end-diastolic volume (RVEDV and LVEDV) measurements another day. Mean LVEDV was comparable between CMR and CARTO (328 +/- 95 and 320 +/- 92 mL, respectively; P = NS), whereas RV volumes measured by CARTO were larger (CMR 140 +/- 48 vs. CARTO 176 +/- 47 mL; P < 0.01). Overall, we found a good correlation between CMR and CARTO measurements for both chambers; however, the Bland-Altman analysis showed a non-interchangeability of these methods. Measurement differences were independent of chamber size, but significantly affected by the number of acquired mapping points. CONCLUSION: Although CMR and CARTO showed a good correlation in the measurement of RVEDV and LVEDV in a group of heart failure patients, the clinical interchangeability of the two methods may be questioned.
No takes yet. Share an insight, caveat, or question.
Grothues et al. (2005) conducted a cross-sectional in Heart failure (NYHA class III/IV) (n=34). CARTO electrophysiological mapping system vs. Cardiac magnetic resonance (CMR) was evaluated on Left ventricular end-diastolic volume (LVEDV) (p=NS). The CARTO system yielded comparable left ventricular volumes to CMR but larger right ventricular volumes, questioning their clinical interchangeability.